Compatibility and Bias Between Ultrasound-guided Attenuation Parameter and Controlled Attenuation Parameter for Ultrasound-based Hepatic Steatosis Assessment

This retrospective observational study aimed to develop a regression model to estimate the controlled attenuation parameter (CAP) from the ultrasound-guided attenuation parameter (UGAP) attenuation coefficient (AC), and to evaluate the agreement, biases, and potential interchangeability between the 2 measurements in a routine clinical setting. A total of 85 patients who underwent both UGAP and CAP measurements on the same day were included. UGAP-AC values were converted from dB/cm/MHz to dB/m to match CAP units. Pearson's correlation and linear regression analyses were performed to derive a predictive equation for CAP based on UGAP-AC. Agreement and biases between the 2 attenuation values were assessed using Bland-Altman analysis. In addition, repeated-measures analysis of variance was conducted to assess the influence of body mass index (BMI) on measurement differences. UGAP-AC and CAP showed a strong positive correlation (r=0.947, P<0.0001). The regression equation was CAP (dB/m)=371.5× UGAP-AC + 12.3 (R²=0.8946). Bland-Altman analysis revealed a fixed bias of 28.6 dB/m and a BMI-dependent proportional bias. However, this proportional bias disappeared observed when analysis was limited to the clinically relevant CAP range (228 to 300 dB/m), suggesting good comparability between UGAP and CAP within the diagnostic range. UGAP-AC can reliably estimate CAP using a simple linear regression model. While some systematic biases exist, their clinical impact is limited. These findings support the potential interchangeability of UGAP and CAP for noninvasive hepatic steatosis assessment in high-throughput clinical practice.

Authors

Institutions

Publication Details

Journal
Ultrasound Quarterly
Published
2026-01-29
DOI
https://doi.org/10.1097/ruq.0000000000000730
Primary Topic
Liver Disease Diagnosis and Treatment
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Compatibility and Bias Between Ultrasound-guided Attenuation Parameter and Controlled Attenuation Parameter for Ultrasound-based Hepatic Steatosis Assessment

Tomoko Sakata, Rino Hiwatashi, Daisuke Miura
Ultrasound Quarterly
Liver Disease Diagnosis and Treatment
article

Compatibility and Bias Between Ultrasound-guided Attenuation Parameter and Controlled Attenuation Parameter for Ultrasound-based Hepatic Steatosis Assessment

Tomoko Sakata, Rino Hiwatashi, Daisuke Miura
article en

Abstract

This retrospective observational study aimed to develop a regression model to estimate the controlled attenuation parameter (CAP) from the ultrasound-guided attenuation parameter (UGAP) attenuation coefficient (AC), and to evaluate the agreement, biases, and potential interchangeability between the 2 measurements in a routine clinical setting. A total of 85 patients who underwent both UGAP and CAP measurements on the same day were included. UGAP-AC values were converted from dB/cm/MHz to dB/m to match CAP units. Pearson's correlation and linear regression analyses were performed to derive a predictive equation for CAP based on UGAP-AC. Agreement and biases between the 2 attenuation values were assessed using Bland-Altman analysis. In addition, repeated-measures analysis of variance was conducted to assess the influence of body mass index (BMI) on measurement differences. UGAP-AC and CAP showed a strong positive correlation (r=0.947, P<0.0001). The regression equation was CAP (dB/m)=371.5× UGAP-AC + 12.3 (R²=0.8946). Bland-Altman analysis revealed a fixed bias of 28.6 dB/m and a BMI-dependent proportional bias. However, this proportional bias disappeared observed when analysis was limited to the clinically relevant CAP range (228 to 300 dB/m), suggesting good comparability between UGAP and CAP within the diagnostic range. UGAP-AC can reliably estimate CAP using a simple linear regression model. While some systematic biases exist, their clinical impact is limited. These findings support the potential interchangeability of UGAP and CAP for noninvasive hepatic steatosis assessment in high-throughput clinical practice.

Ultrasound QuarterlyVol. 42(1)
Associated Press (US), Fukuoka Tokushukai Hospital (JP), All-Russian State Scientific Research Institute for Control, Standardization and Certification of Veterinary Preparations (RU), The Stables (GB)
Openalex Percentile: Top 72%
Liver Disease Diagnosis and Treatment
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Compatibility and Bias Between Ultrasound-guided Attenuation Parameter and Controlled Attenuation Parameter for Ultrasound-based Hepatic Steatosis Assessment — Tomoko Sakata, Rino Hiwatashi, et al. · Ultrasound Quarterly (2026) | TGRS Research Map | TGRS