Efficacy and safety of a novel ab interno supraciliary HA-Mg biodegradable glaucoma drainage plate implantation in rabbit eyes

The uveoscleral outflow pathway is one of the important pathways for aqueous humor outflow. Implanting ab interno glaucoma drainage devices through this pathway does not require conjunctival filtering bleb formation, thereby avoiding bleb-related complications. However, permanent drainage devices can easily cause damage to the corneal endothelium. We hypothesize that a novel ab interno supraciliary HA-Mg biodegradable glaucoma drainage plate through the uveoscleral pathway can reduce corneal endothelial cell damage, demonstrate an IOP-lowering effect, and form and maintain a physiological aqueous outflow pathway after complete degradation and absorption. Sixteen New Zealand white rabbits were randomly assigned to three groups: HA-Mg drainage plate group (10 right eyes), trabeculectomy group (6 right eyes), and control group (16 left eyes). Results showed that the intraocular pressure (IOP) in the ab interno plate group was significantly lower than in the other two groups within the first 20 weeks after surgery (P < 0.0001). After 21 weeks, the IOP in the ab interno plate group gradually returned to the levels of the other two groups. Within 5 months after surgery, the plate was completely degraded and absorbed, the aqueous humor drainage pathway extended to the supraciliary space at the anterior chamber angle, and a water sac-like gap formed above the ciliary body. At the 6th month postoperatively, the number of corneal endothelial cells in the ab interno supraciliary HA-Mg drainage plate group was 2446.0 ± 104.3, and in the control group was 2391.67 ± 49.6, revealing no statistically significant difference (t = -1.611, P = 0.168). In summary, the HA-Mg biodegradable glaucoma drainage plate placement in rabbits was well fixed in the supraciliary space. After 5 months of implantation, the internal drainage plate was completely absorbed, and the implantation procedure and degradation process did not cause damage to the corneal endothelial cells. Compared with the trabeculectomy group, the ab interno plate group maintained a significantly lower IOP for a longer period in this normotensive rabbit model. Although an aqueous humor drainage channel was formed after degradation, the IOP gradually returned to the levels of the control group.

Authors

Institutions

Publication Details

Journal
Journal of Materials Science Materials in Medicine
Published
2026-01-29
DOI
https://doi.org/10.1007/s10856-026-07008-3
Primary Topic
Glaucoma and retinal disorders
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Efficacy and safety of a novel ab interno supraciliary HA-Mg biodegradable glaucoma drainage plate implantation in rabbit eyes

Junlong Yu, Xiaomin Zhu, Shicui Xu, Minghe Xiao et al.
Journal of Materials Science Materials in Medicine
Glaucoma and retinal disorders
article

Efficacy and safety of a novel ab interno supraciliary HA-Mg biodegradable glaucoma drainage plate implantation in rabbit eyes

Junlong Yu, Xiaomin Zhu, Shicui Xu, Minghe Xiao, Shuang Yuan, Xiaofeng Li, Kevin Feng, Hong Liu, Cindy Hutnik, Yujie Rao, Lin Xie, Yong Wang, Wangdu Luo, Yi Chen
article en

Abstract

The uveoscleral outflow pathway is one of the important pathways for aqueous humor outflow. Implanting ab interno glaucoma drainage devices through this pathway does not require conjunctival filtering bleb formation, thereby avoiding bleb-related complications. However, permanent drainage devices can easily cause damage to the corneal endothelium. We hypothesize that a novel ab interno supraciliary HA-Mg biodegradable glaucoma drainage plate through the uveoscleral pathway can reduce corneal endothelial cell damage, demonstrate an IOP-lowering effect, and form and maintain a physiological aqueous outflow pathway after complete degradation and absorption. Sixteen New Zealand white rabbits were randomly assigned to three groups: HA-Mg drainage plate group (10 right eyes), trabeculectomy group (6 right eyes), and control group (16 left eyes). Results showed that the intraocular pressure (IOP) in the ab interno plate group was significantly lower than in the other two groups within the first 20 weeks after surgery (P < 0.0001). After 21 weeks, the IOP in the ab interno plate group gradually returned to the levels of the other two groups. Within 5 months after surgery, the plate was completely degraded and absorbed, the aqueous humor drainage pathway extended to the supraciliary space at the anterior chamber angle, and a water sac-like gap formed above the ciliary body. At the 6th month postoperatively, the number of corneal endothelial cells in the ab interno supraciliary HA-Mg drainage plate group was 2446.0 ± 104.3, and in the control group was 2391.67 ± 49.6, revealing no statistically significant difference (t = -1.611, P = 0.168). In summary, the HA-Mg biodegradable glaucoma drainage plate placement in rabbits was well fixed in the supraciliary space. After 5 months of implantation, the internal drainage plate was completely absorbed, and the implantation procedure and degradation process did not cause damage to the corneal endothelial cells. Compared with the trabeculectomy group, the ab interno plate group maintained a significantly lower IOP for a longer period in this normotensive rabbit model. Although an aqueous humor drainage channel was formed after degradation, the IOP gradually returned to the levels of the control group.

Journal of Materials Science Materials in MedicineVol. 37(1)
Western University (CA), Army Medical University (CN), Chongqing University (CN), The Affiliated Yongchuan Hospital of Chongqing Medical University (CN), St Joseph's Health Care (CA), Chongqing Medical University (CN)
Clean water and sanitation
Openalex Percentile: Top 74%
Glaucoma and retinal disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.