Chronic In Vivo Biostability and Biocompatibility Evaluation of Polyether-Urethane-Based Balloon Implants for Cardiac Application in a Porcine Model

Polyurethane-based implantable devices (PUIDs) delivered via catheter are increasingly used in structural heart interventions; however, limited in vivo data exist regarding their long-term biostability and biological safety. This study evaluated a balloon-shaped implant made of Pellethane<sup>®</sup>, a polyether-based polyurethane, designed as a three-dimensional intracardiac spacer and deployed via percutaneous femoral vein access. The device was chronically positioned adjacent to the tricuspid valve annulus in seven pigs for 24 weeks. Explanted devices and surrounding tissues were evaluated through material characterization (SEM, GPC, FT-IR, and <sup>1</sup>H-NMR) and histological analysis. SEM and FT-IR confirmed preserved surface morphology and chemical bonds, GPC showed stable molecular weight, and <sup>1</sup>H-NMR revealed intact urethane and ether linkages. Materials characterization revealed no evidence of hydrolytic or oxidative degradation, indicating structural stability of the devices. Histological analysis showed stable device positioning with minimal thrombosis or inflammatory response. Biocompatibility was confirmed via ISO 10993-1:2018 Standard (International Organization for Standardization (ISO): Geneva, Switzerland, 2018), and extractable substances were evaluated under exhaustive extraction conditions specified by ISO 10993-18:2020 (International Organization for Standardization (ISO): Geneva, Switzerland, 2020), with no toxicologically significant findings. These findings support the long-term biostability and biological safety of the PUIDs in dynamic cardiac environments, informing future design criteria for catheter-delivered cardiovascular devices.

Authors

Institutions

Publication Details

Journal
Bioengineering
Published
2026-01-29
DOI
https://doi.org/10.3390/bioengineering13020168
Primary Topic
Cardiac Valve Diseases and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Chronic In Vivo Biostability and Biocompatibility Evaluation of Polyether-Urethane-Based Balloon Implants for Cardiac Application in a Porcine Model

June H. Kim, Jin-Chang Kim, Jae-Young Seo, Min-ku Chon et al.
Bioengineering
Cardiac Valve Diseases and Treatments
article

Chronic In Vivo Biostability and Biocompatibility Evaluation of Polyether-Urethane-Based Balloon Implants for Cardiac Application in a Porcine Model

June H. Kim, Jin-Chang Kim, Jae-Young Seo, Min-ku Chon, Jun-Yong Park, Kyeong-Deok Song, Hyun-A Song, Min-Gi Kim
article en

Abstract

Polyurethane-based implantable devices (PUIDs) delivered via catheter are increasingly used in structural heart interventions; however, limited in vivo data exist regarding their long-term biostability and biological safety. This study evaluated a balloon-shaped implant made of Pellethane<sup>®</sup>, a polyether-based polyurethane, designed as a three-dimensional intracardiac spacer and deployed via percutaneous femoral vein access. The device was chronically positioned adjacent to the tricuspid valve annulus in seven pigs for 24 weeks. Explanted devices and surrounding tissues were evaluated through material characterization (SEM, GPC, FT-IR, and <sup>1</sup>H-NMR) and histological analysis. SEM and FT-IR confirmed preserved surface morphology and chemical bonds, GPC showed stable molecular weight, and <sup>1</sup>H-NMR revealed intact urethane and ether linkages. Materials characterization revealed no evidence of hydrolytic or oxidative degradation, indicating structural stability of the devices. Histological analysis showed stable device positioning with minimal thrombosis or inflammatory response. Biocompatibility was confirmed via ISO 10993-1:2018 Standard (International Organization for Standardization (ISO): Geneva, Switzerland, 2018), and extractable substances were evaluated under exhaustive extraction conditions specified by ISO 10993-18:2020 (International Organization for Standardization (ISO): Geneva, Switzerland, 2020), with no toxicologically significant findings. These findings support the long-term biostability and biological safety of the PUIDs in dynamic cardiac environments, informing future design criteria for catheter-delivered cardiovascular devices.

BioengineeringVol. 13(2)
Busan Medical Center (KR), Pusan National University Yangsan Hospital (KR), Pusan National University (KR)
Partnerships for the goals
Openalex Percentile: Top 75%
Cardiac Valve Diseases and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.