Deep-learning-based spectral motion artifact correction on photon-counting cardiac CT images

<i>Objective.</i>While photon-counting computed tomography (PCCT) improves image quality and reduces radiation dose, artifacts induced by cardiac and respiratory motion is still a challenge. The purpose of this work is to evaluate the potential of an image-domain motion-artifact-correction method based on a deep-learning model that incorporates spectral information (material basis images).<i>Approach.</i>We simulated PCCT imaging of five XCAT phantoms, and used these for training two deep neural networks-one with and one without spectral information-to map two motion-corrupted virtual monoenergetic images to corresponding motion-free images. Using images from another simulated XCAT phantom, we calculated the CT number error on five regions of interest and 10 segmented organs. The method was also evaluated visually on clinical cardiac PCCT images. Stretch quantification of endocardial engraved zones was used to calculate regional wall motion and mechanical delay. The results were compared with the motion-free image using a paired t-test.<i>Main results.</i>Out of 45 regions and organs, the CT number accuracy is improved in 41 regions (91%). Among these, the best accuracy is obtained with spectral information in 25 regions (61%). Both models, in particular the one with spectral information, improves visual image quality in simulated and clinical images. The model significantly (<i>P</i>< 0.01) improved estimation of the regional wall motion and assessment of mechanical delay of the left ventricle, but no significant difference was observed between models with and without spectral information.<i>Significance.</i>Our approach, validated on simulated datasets, shows that quantitative cardiac CT imaging can be improved by deep-learning motion correction and that spectral information substantially improves performance.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2026-01-28
DOI
https://doi.org/10.1088/1361-6560/ae3eee
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Deep-learning-based spectral motion artifact correction on photon-counting cardiac CT images

Thomas W. Holmes, Staffan Holmin, Ruihan Huang, Amir Pourmorteza et al.
Physics in Medicine and Biology
Advanced X-ray and CT Imaging
article

Deep-learning-based spectral motion artifact correction on photon-counting cardiac CT images

Thomas W. Holmes, Staffan Holmin, Ruihan Huang, Amir Pourmorteza, Mats Persson, Dennis Hein, Karin Larsson
article en

Abstract

<i>Objective.</i>While photon-counting computed tomography (PCCT) improves image quality and reduces radiation dose, artifacts induced by cardiac and respiratory motion is still a challenge. The purpose of this work is to evaluate the potential of an image-domain motion-artifact-correction method based on a deep-learning model that incorporates spectral information (material basis images).<i>Approach.</i>We simulated PCCT imaging of five XCAT phantoms, and used these for training two deep neural networks-one with and one without spectral information-to map two motion-corrupted virtual monoenergetic images to corresponding motion-free images. Using images from another simulated XCAT phantom, we calculated the CT number error on five regions of interest and 10 segmented organs. The method was also evaluated visually on clinical cardiac PCCT images. Stretch quantification of endocardial engraved zones was used to calculate regional wall motion and mechanical delay. The results were compared with the motion-free image using a paired t-test.<i>Main results.</i>Out of 45 regions and organs, the CT number accuracy is improved in 41 regions (91%). Among these, the best accuracy is obtained with spectral information in 25 regions (61%). Both models, in particular the one with spectral information, improves visual image quality in simulated and clinical images. The model significantly (<i>P</i>< 0.01) improved estimation of the regional wall motion and assessment of mechanical delay of the left ventricle, but no significant difference was observed between models with and without spectral information.<i>Significance.</i>Our approach, validated on simulated datasets, shows that quantitative cardiac CT imaging can be improved by deep-learning motion correction and that spectral information substantially improves performance.

Physics in Medicine and BiologyVol. 71(5)
Georgia Institute of Technology (US), Karolinska University Hospital (SE), Emory University (US), Karolinska Institutet (SE), KTH Royal Institute of Technology (SE)
Openalex Percentile: Top 75%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.