Adaptive optimization framework for accurate multi-material decomposition in dual-energy CT

<i>Purpose</i>. This study introduces an Adaptive Optimization-based Multi-Material Decomposition with Total Nuclear Variation (AO-MMD-TNV) for dual-energy CT (DECT), designed to achieve accurate and noise-robust material fraction estimation.<i>Methods</i>. Three experimental datasets-a digital phantom, a tissue characterization phantom, and a human-shaped phantom-were used to evaluate the proposed method. The algorithm combines a robust Huber data term, iteratively reweighted L1 sparsity, and Total Nuclear Variation (TNV) regularization under simplex constraints. Adaptive weighting was applied to balance physical fidelity, sparsity, and boundary coherence. The performance was quantitatively compared with a conventional triplet-based MMD approach using metrics of volume fraction accuracy (VFA) and standard deviation (STD).<i>Results</i>. The AO-MMD-TNV achieved 100% VFA for all five materials in the digital phantom, with complete noise suppression (STD = 0). In the tissue characterization phantom, the method demonstrated improved VFA in most ROIs, though minor accuracy reductions were observed for breast and solid water (ROIs 3 and 4) due to their similar attenuation to background. In the human-shaped phantom, the AO-MMD-TNV maintained stable performance across six organ-related ROIs, outperforming MMD in quantitative consistency and noise robustness.<i>Conclusions</i>. The proposed AO-MMD-TNV framework effectively enhances quantitative reliability, reduces noise and crosstalk, and preserves anatomical boundaries. While further optimization is needed for low-contrast materials, AO-MMD-TNV demonstrates strong potential for precise, physically consistent DECT-based material quantification and clinical application.

Authors

Institutions

Publication Details

Journal
Biomedical Physics & Engineering Express
Published
2026-01-28
DOI
https://doi.org/10.1088/2057-1976/ae3e99
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Adaptive optimization framework for accurate multi-material decomposition in dual-energy CT

Myung-Ae Chung, Daehong Kim, Hyo-Bin Lee, Haenghwa Lee
Biomedical Physics & Engineering Express
Advanced X-ray and CT Imaging
article

Adaptive optimization framework for accurate multi-material decomposition in dual-energy CT

Myung-Ae Chung, Daehong Kim, Hyo-Bin Lee, Haenghwa Lee
article en

Abstract

<i>Purpose</i>. This study introduces an Adaptive Optimization-based Multi-Material Decomposition with Total Nuclear Variation (AO-MMD-TNV) for dual-energy CT (DECT), designed to achieve accurate and noise-robust material fraction estimation.<i>Methods</i>. Three experimental datasets-a digital phantom, a tissue characterization phantom, and a human-shaped phantom-were used to evaluate the proposed method. The algorithm combines a robust Huber data term, iteratively reweighted L1 sparsity, and Total Nuclear Variation (TNV) regularization under simplex constraints. Adaptive weighting was applied to balance physical fidelity, sparsity, and boundary coherence. The performance was quantitatively compared with a conventional triplet-based MMD approach using metrics of volume fraction accuracy (VFA) and standard deviation (STD).<i>Results</i>. The AO-MMD-TNV achieved 100% VFA for all five materials in the digital phantom, with complete noise suppression (STD = 0). In the tissue characterization phantom, the method demonstrated improved VFA in most ROIs, though minor accuracy reductions were observed for breast and solid water (ROIs 3 and 4) due to their similar attenuation to background. In the human-shaped phantom, the AO-MMD-TNV maintained stable performance across six organ-related ROIs, outperforming MMD in quantitative consistency and noise robustness.<i>Conclusions</i>. The proposed AO-MMD-TNV framework effectively enhances quantitative reliability, reduces noise and crosstalk, and preserves anatomical boundaries. While further optimization is needed for low-contrast materials, AO-MMD-TNV demonstrates strong potential for precise, physically consistent DECT-based material quantification and clinical application.

Biomedical Physics & Engineering ExpressVol. 12(1)
Eulji University (KR), National Cancer Center (KR)
Affordable and clean energy
Openalex Percentile: Top 75%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.