Detector Design using Block Scintillator with 4-Sided Light Sensor and MLPE for Achieving High Resolution and High Sensitivity in Small Animal PET/MRI

A detector utilizing a block scintillator and a semiconductor light sensor was designed for small animal positron emission tomography/magnetic resonance imaging (PET/MRI) applications that achieves high resolution and sensitivity without signal distortion in a magnetic field. The goal was to achieve high sensitivity by using a block scintillator rather than a pixelated scintillator, and high resolution by applying a maximum likelihood position estimation (MLPE). A DETECT2000 simulation was performed to evaluate the performance of the designed detector. Gamma-ray events, where the scintillator and gamma-rays interact, were generated at 1 mm intervals in all directions within the scintillator, and signals were collected by the light sensor. A look-up table (LUT) was created based on the collected signals, and the measurement accuracy of the gamma-ray interaction location was evaluated using the LUT and MLPE. The results showed an excellent measurement accuracy of approximately 85.7% on average. It is considered that this detector can achieve excellent sensitivity and resolution when used in small animal PET/MRI.

Authors

Publication Details

Journal
Journal of Magnetics
Published
2025-12-31
DOI
https://doi.org/10.4283/jmag.2025.30.4.746
Primary Topic
Radiation Detection and Scintillator Technologies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Detector Design using Block Scintillator with 4-Sided Light Sensor and MLPE for Achieving High Resolution and High Sensitivity in Small Animal PET/MRI

Seung-Jae Lee
Journal of Magnetics
Radiation Detection and Scintillator Technologies
article

Detector Design using Block Scintillator with 4-Sided Light Sensor and MLPE for Achieving High Resolution and High Sensitivity in Small Animal PET/MRI

Seung-Jae Lee
article en

Abstract

A detector utilizing a block scintillator and a semiconductor light sensor was designed for small animal positron emission tomography/magnetic resonance imaging (PET/MRI) applications that achieves high resolution and sensitivity without signal distortion in a magnetic field. The goal was to achieve high sensitivity by using a block scintillator rather than a pixelated scintillator, and high resolution by applying a maximum likelihood position estimation (MLPE). A DETECT2000 simulation was performed to evaluate the performance of the designed detector. Gamma-ray events, where the scintillator and gamma-rays interact, were generated at 1 mm intervals in all directions within the scintillator, and signals were collected by the light sensor. A look-up table (LUT) was created based on the collected signals, and the measurement accuracy of the gamma-ray interaction location was evaluated using the LUT and MLPE. The results showed an excellent measurement accuracy of approximately 85.7% on average. It is considered that this detector can achieve excellent sensitivity and resolution when used in small animal PET/MRI.

Journal of MagneticsVol. 30(4)
Openalex Percentile: Top 24%
Radiation Detection and Scintillator Technologies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.