Validation of the automated oscillometric upper-arm cuff home blood pressure monitor Prolife PA5 Comfort AF (BP 136A) in a general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension

This study evaluated the accuracy of the automated oscillometric upper-arm cuff blood pressure (BP) measuring device Prolife PA5 Comfort AF (BP 136A) for home use in a general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2 : 2018) and its amendments (1 : 2020 and 2 : 2024). The requirements of the Universal Standardand its amendments for a validation study in a general population using the same arm sequential measurement method were applied. Participants were recruited to fulfill the age, sex, BP level, arm circumference, and cuff size distribution criteria. The test device measures BP during inflation and uses a wide-range cuff for arm circumference 22-42 cm. One-hundred-twenty-eight individuals were recruited and 85 were analyzed [mean age: 54.9 ± 15.0 (SD) years, 36 (42%) men, arm circumference: 31.7 ± 5.4 cm, range: 22.5-41.9 cm). For validation Criterion 1, the mean difference ± SD between the test device and reference BP readings (N = 255) was -2.1 ± 7.5/-0.3 ± 5.3 mmHg (systolic/diastolic; threshold: ≤5 ± 8 mmHg). For Criterion 2, the SD of the averaged BP differences between the test device and reference BP per individual (N = 85) were 6.44/5.07 mmHg (systolic/diastolic; threshold: ≤6.62/6.95 mmHg). The automated oscillometric home BP monitor Prolife PA5 Comfort AF (BP 136A) fulfills all the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-2 : 2018) and its amendments (1 : 2020 and 2 : 2024) for a general population and can be recommended for clinical use.

Authors

Institutions

Publication Details

Journal
Blood Pressure Monitoring
Published
2026-01-26
DOI
https://doi.org/10.1097/mbp.0000000000000793
Primary Topic
Blood Pressure and Hypertension Studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Validation of the automated oscillometric upper-arm cuff home blood pressure monitor Prolife PA5 Comfort AF (BP 136A) in a general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension

Vasileios Ntousopoulos, Konstantina-Rozalia Bakalakou, Petroula Lempidaki, Evangelia Ntaskagianni et al.
Blood Pressure Monitoring
Blood Pressure and Hypertension Studies
article

Validation of the automated oscillometric upper-arm cuff home blood pressure monitor Prolife PA5 Comfort AF (BP 136A) in a general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension

Vasileios Ntousopoulos, Konstantina-Rozalia Bakalakou, Petroula Lempidaki, Evangelia Ntaskagianni, George S. Stergiou, Anastasios Kollias, Aikaterini Komnianou, Dimitrios Mariglis, Ariadni Menti
article en

Abstract

This study evaluated the accuracy of the automated oscillometric upper-arm cuff blood pressure (BP) measuring device Prolife PA5 Comfort AF (BP 136A) for home use in a general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2 : 2018) and its amendments (1 : 2020 and 2 : 2024). The requirements of the Universal Standardand its amendments for a validation study in a general population using the same arm sequential measurement method were applied. Participants were recruited to fulfill the age, sex, BP level, arm circumference, and cuff size distribution criteria. The test device measures BP during inflation and uses a wide-range cuff for arm circumference 22-42 cm. One-hundred-twenty-eight individuals were recruited and 85 were analyzed [mean age: 54.9 ± 15.0 (SD) years, 36 (42%) men, arm circumference: 31.7 ± 5.4 cm, range: 22.5-41.9 cm). For validation Criterion 1, the mean difference ± SD between the test device and reference BP readings (N = 255) was -2.1 ± 7.5/-0.3 ± 5.3 mmHg (systolic/diastolic; threshold: ≤5 ± 8 mmHg). For Criterion 2, the SD of the averaged BP differences between the test device and reference BP per individual (N = 85) were 6.44/5.07 mmHg (systolic/diastolic; threshold: ≤6.62/6.95 mmHg). The automated oscillometric home BP monitor Prolife PA5 Comfort AF (BP 136A) fulfills all the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-2 : 2018) and its amendments (1 : 2020 and 2 : 2024) for a general population and can be recommended for clinical use.

Blood Pressure Monitoring
National and Kapodistrian University of Athens (GR), Sotiria General Hospital (GR)
Partnerships for the goals
Openalex Percentile: Top 79%
Blood Pressure and Hypertension Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.