Phantom-based performance comparison of two commercial deep learning CT reconstruction algorithms with super- and normal-resolution settings
The performance of a new deep learning super-resolution image reconstruction algorithm (SR-DLR) was compared to a normal-resolution (NR)-DLR algorithm using an image-quality phantom for an abdominal energy-integrating detector CT protocol. SR-DLR with a 1,024<sup>2</sup> matrix improved spatial resolution and detectability of simulated abdominal lesions compared to NR-DLR with a 512<sup>2</sup> matrix. Using SR-DLR, therefore, presents numerous prospects for improving abdominal CT images and a high potential for reducing the radiation doses.