Asymptomatic central venous occlusion secondary to central venous catheter-use complicating pacemaker implantation: a leadless solution
Abnormal central venous anatomy can obstruct cardiac-device implantation. We report a 54-year-old patient found to be in complete heart-block after identification of profound bradycardia, requiring cardiac-pacing. Echocardiography revealed systolic dysfunction, prompting a plan for cardiac resynchronization therapy. Conventional lead placement was impossible due to failed guide-wire advancement bilaterally. Peri-procedural venography demonstrated attenuation of both brachiocephalic veins and contrast-enhanced computed tomography confirmed complete central venous occlusion (CVO). Further history identified previous chemotherapy delivered via central venous catheters, as the likely aetiology for the CVO <i>and</i> systolic dysfunction. Alternative pacing options were considered. Ultimately the leadless <i>Micra-AV</i>-pacemaker was successfully implanted via the femoral vein with good clinical recovery. This unique case highlights CVO as an obstacle to cardiac-device implantation, with specific patients at increased risk. Pre-procedural imaging in those at-risk may reduce procedure failure and facilitate appropriate device strategy choice. Leadless pacing provides a safe, effective alternative in cases of CVO.
Authors
- Christopher Cassidy (ORCID: https://orcid.org/0000-0002-6730-4039)
- Kanarath Balachandran
- James Baudry
Institutions
- University of Liverpool (GB)
- Blackpool Victoria Hospital (GB)
- Royal Blackburn Teaching Hospital (GB)
- East Lancashire Hospitals NHS Trust (GB)
Publication Details
- Journal
- Oxford Medical Case Reports
- Published
- 2025-12-11
- DOI
- https://doi.org/10.1093/omcr/omaf292
- Primary Topic
- Central Venous Catheters and Hemodialysis
- Type
- article
- Field-Weighted Citation Impact
- 0.00