Chest Radiography Optimization: Identifying the Optimal kV for Image Quality in a Phantom Study

Chest radiography remains one of the most frequently performed imaging examinations, highlighting the need for optimization of acquisition parameters to balance image quality and radiation dose. This study presents a phantom-based quantitative evaluation of chest radiography acquisition settings using a digital radiography system (AGFA DR 600). Measurements were performed at three tube voltage levels across simulated patient-equivalent thicknesses generated using PMMA slabs, with a Leeds TOR 15FG image quality phantom positioned centrally in the imaging setup. Image quality was quantitatively assessed using signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR), which were calculated from mean pixel values obtained from repeated acquisitions. Radiation exposure was evaluated through estimation of entrance surface dose (ESD). The analysis demonstrated that dose-normalized performance metrics favored intermediate tube voltages for slim and average patient-equivalent thicknesses, while higher voltages were required to maintain image quality in obese-equivalent conditions. Overall, image quality and dose were found to be strongly dependent on the combined selection of tube voltage and phantom thickness. These findings indicate that modest adjustments to tube voltage selection may improve the balance between image quality and radiation dose in chest radiography. Nevertheless, as the present work is based on phantom measurements, further validation using clinical images and observer-based studies is required before any modification of routine radiographic practice.

Authors

Institutions

Publication Details

Journal
Journal of Imaging
Published
2026-01-21
DOI
https://doi.org/10.3390/jimaging12010049
Primary Topic
Digital Radiography and Breast Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Chest Radiography Optimization: Identifying the Optimal kV for Image Quality in a Phantom Study

Kyriakos Kokkinogoulis, Maria Giannopoulou, Efstathios P. Efstathopoulos, Ioannis Antonakos
Journal of Imaging
Digital Radiography and Breast Imaging
article

Chest Radiography Optimization: Identifying the Optimal kV for Image Quality in a Phantom Study

Kyriakos Kokkinogoulis, Maria Giannopoulou, Efstathios P. Efstathopoulos, Ioannis Antonakos
article en

Abstract

Chest radiography remains one of the most frequently performed imaging examinations, highlighting the need for optimization of acquisition parameters to balance image quality and radiation dose. This study presents a phantom-based quantitative evaluation of chest radiography acquisition settings using a digital radiography system (AGFA DR 600). Measurements were performed at three tube voltage levels across simulated patient-equivalent thicknesses generated using PMMA slabs, with a Leeds TOR 15FG image quality phantom positioned centrally in the imaging setup. Image quality was quantitatively assessed using signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR), which were calculated from mean pixel values obtained from repeated acquisitions. Radiation exposure was evaluated through estimation of entrance surface dose (ESD). The analysis demonstrated that dose-normalized performance metrics favored intermediate tube voltages for slim and average patient-equivalent thicknesses, while higher voltages were required to maintain image quality in obese-equivalent conditions. Overall, image quality and dose were found to be strongly dependent on the combined selection of tube voltage and phantom thickness. These findings indicate that modest adjustments to tube voltage selection may improve the balance between image quality and radiation dose in chest radiography. Nevertheless, as the present work is based on phantom measurements, further validation using clinical images and observer-based studies is required before any modification of routine radiographic practice.

Journal of ImagingVol. 12(1)
National and Kapodistrian University of Athens (GR)
Openalex Percentile: Top 83%
Digital Radiography and Breast Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.