Local Diagnostic Reference Levels for Intracranial Aneurysm Coil-Only Embolization Using a Low-Dose Technique
<b>Background/Objectives:</b> Optimizing routine neurointerventional workflow and minimizing exposure to ionizing radiation during coil-only endovascular treatment of intracranial aneurysms depend on operator experience, reduced frame rates during both fluoroscopy and digital subtraction angiography (DSA), and the use of advanced angiographic systems. The low-dose protocol implemented in this study used the lowest available fluoroscopy frame rate (3.125 frames per second [fps]) and a nominal acquisition rate of 2 fps (actual = 2.45 fps) for DSA, three-dimensional (3D) rotational angiography, two-dimensional (2D)/3D mapping, and roadmapping. <b>Methods:</b> This retrospective analysis encompassed 245 coil-only procedures performed at a single tertiary center from 2018 to 2024. Data collected for each procedure included dose-area product (DAP), reference air kerma (K<sub>a,r</sub>), fluoroscopy time (FT), and the total number of DSA frames. Local diagnostic reference levels (DRLs; 75th percentile [P75]) and typical values (50th percentile [P50]) were determined and descriptively compared with values reported in the literature. <b>Results:</b> The P75 values, representing DRLs, were 22.4 Gy·cm<sup>2</sup> for DAP (literature range, 123-272.8 Gy·cm<sup>2</sup>), 268 mGy for K<sub>a,r</sub> (1171-4240 mGy), 18 min 56 s for FT, and 285 DSA frames. The P50 values were 13.8 Gy·cm<sup>2</sup> for DAP (78.7-179.0 Gy·cm<sup>2</sup>), 196 mGy for K<sub>a,r</sub> (801-2804 mGy), 13 min 25 s for FT, and 208 DSA frames. <b>Conclusions:</b> In this single-center cohort, dose metrics for coil-only intracranial aneurysm treatment were within the lower range of published values. Cross-study comparisons are descriptive and require cautious interpretation. The proposed local DRLs may support quality assurance, dose optimization, and patient safety in comparable clinical settings. Further multi-center and multi-operator studies are warranted to evaluate transferability and applicability beyond coil-only procedures.
Authors
- M. Budzanowski (ORCID: https://orcid.org/0000-0001-9426-7619)
- Mariusz Sowa
- Kamil Węglarz
- Joanna Sowa (ORCID: https://orcid.org/0000-0002-2402-5363)
Institutions
- Institute of Nuclear Physics, Polish Academy of Sciences (PL)
- University of Computer Sciences and Economics in Olsztyn (PL)
- Polish Academy of Learning (PL)
- University of Warmia and Mazury in Olsztyn (PL)
Publication Details
- Journal
- Biomedicines
- Published
- 2026-01-21
- DOI
- https://doi.org/10.3390/biomedicines14010233
- Primary Topic
- Radiation Dose and Imaging
- Type
- article
- Field-Weighted Citation Impact
- 0.00