Erratic Pulse Oximetry Pulse Rate With Preserved Oxygen Saturation in a Patient Wearing White Nail Polish: A Case Report

Pulse oximetry is widely used for noninvasive monitoring of oxygen saturation (SpO₂) and pulse rate. Because the pulse rate displayed by a pulse oximeter is derived from the peripheral photoplethysmographic waveform, it can be affected by artifact even when oxygen saturation appears normal. We report a case of an asymptomatic patient wearing white nail polish who demonstrated stable, normal SpO₂ values yet markedly erratic pulse rate readings on pulse oximetry, fluctuating rapidly between bradycardic and tachycardic ranges. Concern for arrhythmia prompted a 12‑lead electrocardiogram, which showed a normal sinus rhythm. This report highlights a practical diagnostic pitfall: nail cosmetics may have little effect on SpO₂ yet still destabilize pulse waveform detection and generate misleading pulse rate values, leading to unnecessary evaluation and erroneous clinical conclusions.

Authors

Institutions

Publication Details

Journal
Cureus
Published
2026-01-21
DOI
https://doi.org/10.7759/cureus.101995
Primary Topic
Non-Invasive Vital Sign Monitoring
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Erratic Pulse Oximetry Pulse Rate With Preserved Oxygen Saturation in a Patient Wearing White Nail Polish: A Case Report

Akhtar Purvez
Cureus
Non-Invasive Vital Sign Monitoring
article

Erratic Pulse Oximetry Pulse Rate With Preserved Oxygen Saturation in a Patient Wearing White Nail Polish: A Case Report

Akhtar Purvez
article en

Abstract

Pulse oximetry is widely used for noninvasive monitoring of oxygen saturation (SpO₂) and pulse rate. Because the pulse rate displayed by a pulse oximeter is derived from the peripheral photoplethysmographic waveform, it can be affected by artifact even when oxygen saturation appears normal. We report a case of an asymptomatic patient wearing white nail polish who demonstrated stable, normal SpO₂ values yet markedly erratic pulse rate readings on pulse oximetry, fluctuating rapidly between bradycardic and tachycardic ranges. Concern for arrhythmia prompted a 12‑lead electrocardiogram, which showed a normal sinus rhythm. This report highlights a practical diagnostic pitfall: nail cosmetics may have little effect on SpO₂ yet still destabilize pulse waveform detection and generate misleading pulse rate values, leading to unnecessary evaluation and erroneous clinical conclusions.

CureusVol. 18(1)
Lincoln Memorial University (US), Charlottesville Medical Research (US), Momentum Research (US)
Quality Education
Openalex Percentile: Top 83%
Non-Invasive Vital Sign Monitoring
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.