Upgraded beam profile monitoring using Chromox and fiber optic imaging for high-radiation environments

We present an upgraded beam monitoring system designed for use in high-radiation environments where conventional imaging solutions rapidly degrade. In the presented device the radiation-sensitive P47 phosphor screen of the previous system* is replaced with a radiation-hard Chromox ceramic scintillator and relocates the CMOS camera outside the irradiation zone by transmitting the optical signal through a 20-meter radiation-tolerant fiber optic bundle. To enhance operational flexibility, a pneumatic actuator enables remote insertion and retraction of the scintillating screen into the beam path. The radiation hardness of the Chromox ceramic and the fiber bundle was evaluated, and the optical system’s resolution and fidelity were characterized. The new setup demonstrated stable imaging performance under irradiation, addressing the frequent maintenance issues of the previous system. These improvements offer a robust and low-maintenance solution for beam profile monitoring in demanding accelerator environments.

Authors

Publication Details

Journal
JACOW
Published
2026-01-20
DOI
https://doi.org/10.18429/jacow-ibic2025-tupmo16
Citations
1
Primary Topic
Radiation Detection and Scintillator Technologies
Type
article
Field-Weighted Citation Impact
47.32
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Upgraded beam profile monitoring using Chromox and fiber optic imaging for high-radiation environments

Alexander Gottstein, Isidre Mateu, Saverio Braccini, Eva Kasanda et al.
1 citations
JACOW
Radiation Detection and Scintillator Technologies
47.32
article

Upgraded beam profile monitoring using Chromox and fiber optic imaging for high-radiation environments

Alexander Gottstein, Isidre Mateu, Saverio Braccini, Eva Kasanda, Lars Eggimann
article en
1 citations

Abstract

We present an upgraded beam monitoring system designed for use in high-radiation environments where conventional imaging solutions rapidly degrade. In the presented device the radiation-sensitive P47 phosphor screen of the previous system* is replaced with a radiation-hard Chromox ceramic scintillator and relocates the CMOS camera outside the irradiation zone by transmitting the optical signal through a 20-meter radiation-tolerant fiber optic bundle. To enhance operational flexibility, a pneumatic actuator enables remote insertion and retraction of the scintillating screen into the beam path. The radiation hardness of the Chromox ceramic and the fiber bundle was evaluated, and the optical system’s resolution and fidelity were characterized. The new setup demonstrated stable imaging performance under irradiation, addressing the frequent maintenance issues of the previous system. These improvements offer a robust and low-maintenance solution for beam profile monitoring in demanding accelerator environments.

JACOW
Openalex Percentile: Top 1%
Radiation Detection and Scintillator Technologies
47.32
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.