Measuring and determination of tidal volume of the ventilator tester by different methods

Accurate tidal volume measurement is essential for the calibration and verification of ventilator testers to ensure the reliable performance of mechanical ventilators. This study compared and evaluated three calibration methods: A Molbloc reference flow meter, wet gas reference flow meter, and calibration volumetric syringe. The Molbloc flow meter demonstrated consistent results with a measurement uncertainty of ±0.29%. While the wet gas meter exhibited minimal error at lower volumes but showed increased inaccuracies at a volume of 400 mL due to rotor movement issues. Conversely, the volumetric syringe method provided the highest accuracy, with a measurement uncertainty of ±0.27%. The high accuracy is mostly due to its direct volumetric displacement and precise calibration standards. The syringe method also simulated real human breathing dynamics across a wide range of flow rates (0.1–34 L/min), enhancing its applicability in realistic ventilator testing. Comparative analysis revealed that the syringe method is the most reliable and precise, making it the preferred choice for tidal volume calibration in clinical and laboratory settings.

Authors

Institutions

Publication Details

Journal
Proceedings of the Institution of Mechanical Engineers Part C Journal of Mechanical Engineering Science
Published
2026-01-19
DOI
https://doi.org/10.1177/09544062251410252
Primary Topic
Respiratory Support and Mechanisms
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Measuring and determination of tidal volume of the ventilator tester by different methods

Ali Zahran, Ali E. Abuelezz, Fatma A.-M. Kassem
Proceedings of the Institution of Mechanical Engineers Part C Journal of Mechanical Engineering Science
Respiratory Support and Mechanisms
article

Measuring and determination of tidal volume of the ventilator tester by different methods

Ali Zahran, Ali E. Abuelezz, Fatma A.-M. Kassem
article en

Abstract

Accurate tidal volume measurement is essential for the calibration and verification of ventilator testers to ensure the reliable performance of mechanical ventilators. This study compared and evaluated three calibration methods: A Molbloc reference flow meter, wet gas reference flow meter, and calibration volumetric syringe. The Molbloc flow meter demonstrated consistent results with a measurement uncertainty of ±0.29%. While the wet gas meter exhibited minimal error at lower volumes but showed increased inaccuracies at a volume of 400 mL due to rotor movement issues. Conversely, the volumetric syringe method provided the highest accuracy, with a measurement uncertainty of ±0.27%. The high accuracy is mostly due to its direct volumetric displacement and precise calibration standards. The syringe method also simulated real human breathing dynamics across a wide range of flow rates (0.1–34 L/min), enhancing its applicability in realistic ventilator testing. Comparative analysis revealed that the syringe method is the most reliable and precise, making it the preferred choice for tidal volume calibration in clinical and laboratory settings.

Proceedings of the Institution of Mechanical Engineers Part C Journal of Mechanical Engineering ScienceVol. 240(7)
National Institute of Standards (EG)
Openalex Percentile: Top 85%
Respiratory Support and Mechanisms
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.