Spectroradiometric calibration and temperature sensitivity analysis of a fluorescence box

The fluorescence box (FloX) is a hyperspectral instrument designed for unattended observation of sun-induced chlorophyll fluorescence (SIF) and reflectance. FloX instruments are being used as part of a validation network for the upcoming Fluorescence Explorer imaging satellite. Hence, these devices are expected to deliver continuous observations with high accuracy. We perform the spectroradiometric calibration and temperature sensitivity analysis of a FloX unit, which comprises two spectrometers. We present the results of a thermal cycling test of the FloX using a refrigerated incubator, spanning temperatures from 40°C to 0°C in steps of 5°C, during which the instrument’s optical fibers used an access port to carry out external measurements of the reflected radiance from a Spectralon panel illuminated by a tungsten lamp. The mean radiance values within the SIF spectral range exhibited changes of over 3% for the high-resolution sensor and over 4% for the broadband sensor, relative to the nominal radiance values at 20°C. We examine the contributions to the combined standard uncertainty to show how these radiometric shifts are a systematic effect of changing environmental temperature.

Authors

Publication Details

Journal
Universität Zürich, ZORA
Published
2025-08-18
DOI
https://doi.org/10.5167/uzh-282239
Primary Topic
Calibration and Measurement Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Spectroradiometric calibration and temperature sensitivity analysis of a fluorescence box

Simon A. Trim, Andreas Hüni
Universität Zürich, ZORA
Calibration and Measurement Techniques
article

Spectroradiometric calibration and temperature sensitivity analysis of a fluorescence box

Simon A. Trim, Andreas Hüni
article en

Abstract

The fluorescence box (FloX) is a hyperspectral instrument designed for unattended observation of sun-induced chlorophyll fluorescence (SIF) and reflectance. FloX instruments are being used as part of a validation network for the upcoming Fluorescence Explorer imaging satellite. Hence, these devices are expected to deliver continuous observations with high accuracy. We perform the spectroradiometric calibration and temperature sensitivity analysis of a FloX unit, which comprises two spectrometers. We present the results of a thermal cycling test of the FloX using a refrigerated incubator, spanning temperatures from 40°C to 0°C in steps of 5°C, during which the instrument’s optical fibers used an access port to carry out external measurements of the reflected radiance from a Spectralon panel illuminated by a tungsten lamp. The mean radiance values within the SIF spectral range exhibited changes of over 3% for the high-resolution sensor and over 4% for the broadband sensor, relative to the nominal radiance values at 20°C. We examine the contributions to the combined standard uncertainty to show how these radiometric shifts are a systematic effect of changing environmental temperature.

Universität Zürich, ZORA
Openalex Percentile: Top 16%
Calibration and Measurement Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.