Early aseptic loosening and inferior patient-reported outcomes of a cementless tibial baseplate in a modern total knee arthroplasty design
Aims: This study reports the outcome of a modern total knee arthroplasty design using a cementless tibial baseplate compared to the full-cemented version. Methods: Consecutive cohorts with 12-month follow-up were evaluated. Patients receiving a cementless tibial baseplate were compared to those who received a cemented tibial component. Endpoints included revision rates and reason for revision, patient-reported outcome measures (PROMs) using the Oxford Knee Score (OKS), and progressive radiolucency. Retrieval analysis was performed for the revised cases. Pearson correlation analysis and multiple regression analysis were used. Conclusion: In our cohort, there was a significantly higher incidence of aseptic loosening and worse PROMs at one year for the cementless tibial baseplate. An increased BMI may be an independent risk factor for aseptic loosening and inferior PROMs.
Authors
- Octavian Andronic
- Moreica Pabbruwe (ORCID: https://orcid.org/0000-0002-4223-0416)
- Yue H Yang
- Chris W Jones
- Piers J Yates
Publication Details
- Journal
- Universität Zürich, ZORA
- Published
- 2025-04-01
- DOI
- https://doi.org/10.5167/uzh-282329
- Primary Topic
- Total Knee Arthroplasty Outcomes
- Type
- article
- Field-Weighted Citation Impact
- 0.00