Dry eye disease (DED) is a widespread ocular surface disorder that requires early and accurate diagnostics for effective management. Among various biomarkers, matrix metalloproteinase-9 (MMP-9) has been validated as a reliable indicator of ocular inflammation. However, current lateral flow assays (LFAs), such as InflammaDry, provide only binary outcomes, which limits their ability to stratify disease severity and guide personalized treatment. More advanced and accessible diagnostic strategies are needed to address these gaps. This paper presents a portable smartphone-based platform that integrates machine learning (ML) with LFAs for multiclass detection of tear MMP-9. Customized LFA strips were developed to detect six different concentrations of MMP-9, each repeated six times. Images were captured using six smartphones under four distinct backgrounds, generating a dataset of 864 images. The segment anything model (SAM) was applied for strip segmentation, achieving 100% accuracy across all samples. ResNet34 and MobileNet_V2 ML model, pre-trained on ImageNet, were used for classification and achieved accuracies of 99% and over 97%, respectively. The results demonstrate that ML-assisted smartphone imaging enables reliable multiclass detection, overcoming the limitations of conventional binary LFA assays. This approach addresses variability across devices and environments, eliminates the need for external readers, and provides a practical, low-cost solution for DED diagnosis. More broadly, the framework can be extended to other LFA-based assays, supporting scalable point-of-care testing (POCT) in resource-limited settings. • The smartphone-assisted paper-based chemiluminescence detection platform offers high integration. • A fully automated “sample in, answer out” operation fashion. • The chemiluminescence detection process has been optimized for enhanced performance. • Rabbit IgG testing achieved a limit of detection of 62.4 pg/mL. • Detection of tau protein reached a limit of detection of 26.1 pg/mL, which is lower than clinical cut-off value.

Authors

Institutions

Publication Details

Journal
Biomedical Instrumentation
Published
2025-12-01
DOI
https://doi.org/10.1016/j.bmi.2026.100028
Primary Topic
Ocular Surface and Contact Lens
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Portable multiclass lateral flow assay for dry eye disease based on machine learning and smartphone imaging

Ziren Xiao, Sixuan Duan, Kaizhu Huang, Letian Hu et al.
Biomedical Instrumentation
Ocular Surface and Contact Lens
article

Portable multiclass lateral flow assay for dry eye disease based on machine learning and smartphone imaging

Ziren Xiao, Sixuan Duan, Kaizhu Huang, Letian Hu, Pengfei Song, Jihong Sun, Kai Hoettges
article en

Abstract

Dry eye disease (DED) is a widespread ocular surface disorder that requires early and accurate diagnostics for effective management. Among various biomarkers, matrix metalloproteinase-9 (MMP-9) has been validated as a reliable indicator of ocular inflammation. However, current lateral flow assays (LFAs), such as InflammaDry, provide only binary outcomes, which limits their ability to stratify disease severity and guide personalized treatment. More advanced and accessible diagnostic strategies are needed to address these gaps. This paper presents a portable smartphone-based platform that integrates machine learning (ML) with LFAs for multiclass detection of tear MMP-9. Customized LFA strips were developed to detect six different concentrations of MMP-9, each repeated six times. Images were captured using six smartphones under four distinct backgrounds, generating a dataset of 864 images. The segment anything model (SAM) was applied for strip segmentation, achieving 100% accuracy across all samples. ResNet34 and MobileNet_V2 ML model, pre-trained on ImageNet, were used for classification and achieved accuracies of 99% and over 97%, respectively. The results demonstrate that ML-assisted smartphone imaging enables reliable multiclass detection, overcoming the limitations of conventional binary LFA assays. This approach addresses variability across devices and environments, eliminates the need for external readers, and provides a practical, low-cost solution for DED diagnosis. More broadly, the framework can be extended to other LFA-based assays, supporting scalable point-of-care testing (POCT) in resource-limited settings. • The smartphone-assisted paper-based chemiluminescence detection platform offers high integration. • A fully automated “sample in, answer out” operation fashion. • The chemiluminescence detection process has been optimized for enhanced performance. • Rabbit IgG testing achieved a limit of detection of 62.4 pg/mL. • Detection of tau protein reached a limit of detection of 26.1 pg/mL, which is lower than clinical cut-off value.

Biomedical InstrumentationVol. 1(2)
University of Liverpool (GB), Duke Kunshan University (CN), Xi’an Jiaotong-Liverpool University (CN)
Openalex Percentile: Top 23%
Ocular Surface and Contact Lens
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.