Simultaneous Quantification of Dexamethasone and Cortisol for the Dexamethasone Suppression Test Using LC-MS/MS: Adaptation of a Commercial CE-IVD Steroid Assay

Background: The dexamethasone suppression test (DST) is a key diagnostic tool for evaluating disorders of the hypothalamic–pituitary–adrenal (HPA) axis. Interpretation of DST results can be confounded by inadequate dexamethasone exposure due to pharmacokinetic variability or patient nonadherence. Simultaneous measurement of dexamethasone and cortisol by LC-MS/MS improves diagnostic accuracy by distinguishing true biological nonsuppression from insufficient drug exposure and prevents analytical interferences observed with cortisol immunoassays. Methods: A commercially available CE-IVD LC-MS/MS steroid panel (Chromsystems) was adapted for the simultaneous quantification of dexamethasone and cortisol in human plasma. The method was validated for linearity, accuracy, precision, selectivity, carry-over, matrix effects, and stability. Method comparison was performed using 26 clinical DST samples analyzed using immunoassay and external LC-MS testing. The proposed dexamethasone cutoff of 1.3 ng/mL (3.3 nmol/L) was evaluated in 62 patient samples. Results: The method demonstrated excellent linearity, with intra- and interassay accuracy and precision within ±15% for both analytes. No relevant carry-over or matrix effects were observed. Agreement with reference methods was 91% ± 6% for cortisol and 95% ± 7% for dexamethasone. Among the 62 DST samples, 48 showed appropriate suppression [cortisol <18.12 ng/mL (<50 nmol/L)] with dexamethasone levels above the cutoff. In contrast, 3 samples with insufficient suppression [cortisol >48.6 ng/mL (134.1 nmol/L)] had subthreshold dexamethasone levels, whereas 11 samples showed elevated cortisol levels despite sufficient dexamethasone exposure, indicating possible HPA axis dysfunction. Conclusions: The proposed rapid and robust LC-MS/MS method enables reliable, simultaneous quantification of dexamethasone and cortisol. The assay supports accurate DST interpretation by identifying cases of pharmacokinetic variance or nonadherence.

Authors

Publication Details

Journal
Universität Zürich, ZORA
Published
2025-09-30
DOI
https://doi.org/10.5167/uzh-282626
Primary Topic
Adrenal Hormones and Disorders
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Simultaneous Quantification of Dexamethasone and Cortisol for the Dexamethasone Suppression Test Using LC-MS/MS: Adaptation of a Commercial CE-IVD Steroid Assay

Marc Luginbühl, R. Steiner
Universität Zürich, ZORA
Adrenal Hormones and Disorders
article

Simultaneous Quantification of Dexamethasone and Cortisol for the Dexamethasone Suppression Test Using LC-MS/MS: Adaptation of a Commercial CE-IVD Steroid Assay

Marc Luginbühl, R. Steiner
article en

Abstract

Background: The dexamethasone suppression test (DST) is a key diagnostic tool for evaluating disorders of the hypothalamic–pituitary–adrenal (HPA) axis. Interpretation of DST results can be confounded by inadequate dexamethasone exposure due to pharmacokinetic variability or patient nonadherence. Simultaneous measurement of dexamethasone and cortisol by LC-MS/MS improves diagnostic accuracy by distinguishing true biological nonsuppression from insufficient drug exposure and prevents analytical interferences observed with cortisol immunoassays. Methods: A commercially available CE-IVD LC-MS/MS steroid panel (Chromsystems) was adapted for the simultaneous quantification of dexamethasone and cortisol in human plasma. The method was validated for linearity, accuracy, precision, selectivity, carry-over, matrix effects, and stability. Method comparison was performed using 26 clinical DST samples analyzed using immunoassay and external LC-MS testing. The proposed dexamethasone cutoff of 1.3 ng/mL (3.3 nmol/L) was evaluated in 62 patient samples. Results: The method demonstrated excellent linearity, with intra- and interassay accuracy and precision within ±15% for both analytes. No relevant carry-over or matrix effects were observed. Agreement with reference methods was 91% ± 6% for cortisol and 95% ± 7% for dexamethasone. Among the 62 DST samples, 48 showed appropriate suppression [cortisol <18.12 ng/mL (<50 nmol/L)] with dexamethasone levels above the cutoff. In contrast, 3 samples with insufficient suppression [cortisol >48.6 ng/mL (134.1 nmol/L)] had subthreshold dexamethasone levels, whereas 11 samples showed elevated cortisol levels despite sufficient dexamethasone exposure, indicating possible HPA axis dysfunction. Conclusions: The proposed rapid and robust LC-MS/MS method enables reliable, simultaneous quantification of dexamethasone and cortisol. The assay supports accurate DST interpretation by identifying cases of pharmacokinetic variance or nonadherence.

Universität Zürich, ZORA
Good health and well-being
Openalex Percentile: Top 27%
Adrenal Hormones and Disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.