Agreement of respiratory impedance measurements between two commercial forced oscillation technique devices: A randomized crossover study

The performance of forced oscillation technique (FOT) devices may vary due to differences in hardware and signal processing. This study examined the agreement of respiratory impedance (Zrs) measured by two commercial FOT devices in unselected adults with respiratory symptoms. The Zrs of 97 consecutive adults with respiratory symptoms were measured using the TremoFlo C-100 and Resmon Pro FULL V3, in a randomized crossover design. Paired t-tests and Bland–Altman plots with 95% limits of agreement (LoA) were used to compare Zrs measurements between the two devices. The average age of the cohort was 42.3 years, and 46% were women. R5 (p = 0.07) and R19 (p < 0.001) values were higher with Resmon Pro than with the TremoFlo. In contrast, R5–19 was significantly smaller with Resmon Pro, with a paired difference of −0.20 cm H₂O/L/s (p = 0.01). The within- and whole-breath X5 of Resmon Pro were significantly less negative than those of TremoFlo. The reactance area and resonant frequency of Resmon Pro were also significantly smaller than those obtained with TremoFlo. The bias for R5 (0.29 cmH 2 O/L/s; LoA: -2.69, 3.26) was broad and had random errors. The biases for whole-breath X5 (0.8 cmH 2 O/L/s; 95% LoA: -1.25, 2.85) was also wide and AX (-13.66 cmH 2 O/L/s; LoA: -45.46, 18.14) exhibited systematic errors. Zrs measurement differed significantly between the two FOT devices, particularly for reactance parameters. Therefore, normative values derived from one device should be avoided when interpreting results from another. • Respiratory impedance obtained from two oscillometers shows wide random and systematic variability. • Normative equations are device-specific and should not be used to interpret results obtained from other devices. • Standardization across oscillometers is crucial to ensure robust and reliable interpretation

Authors

Institutions

Publication Details

Journal
Pulmonary Function Diagnostics
Published
2026-01-15
DOI
https://doi.org/10.1016/j.pulmfd.2026.100002
Citations
1
Primary Topic
Chronic Obstructive Pulmonary Disease (COPD) Research
Type
article
Field-Weighted Citation Impact
36.44
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Agreement of respiratory impedance measurements between two commercial forced oscillation technique devices: A randomized crossover study

S De
1 citations
Pulmonary Function Diagnostics
Chronic Obstructive Pulmonary Disease (COPD) Research
36.44
article

Agreement of respiratory impedance measurements between two commercial forced oscillation technique devices: A randomized crossover study

S De
article en
1 citations

Abstract

The performance of forced oscillation technique (FOT) devices may vary due to differences in hardware and signal processing. This study examined the agreement of respiratory impedance (Zrs) measured by two commercial FOT devices in unselected adults with respiratory symptoms. The Zrs of 97 consecutive adults with respiratory symptoms were measured using the TremoFlo C-100 and Resmon Pro FULL V3, in a randomized crossover design. Paired t-tests and Bland–Altman plots with 95% limits of agreement (LoA) were used to compare Zrs measurements between the two devices. The average age of the cohort was 42.3 years, and 46% were women. R5 (p = 0.07) and R19 (p < 0.001) values were higher with Resmon Pro than with the TremoFlo. In contrast, R5–19 was significantly smaller with Resmon Pro, with a paired difference of −0.20 cm H₂O/L/s (p = 0.01). The within- and whole-breath X5 of Resmon Pro were significantly less negative than those of TremoFlo. The reactance area and resonant frequency of Resmon Pro were also significantly smaller than those obtained with TremoFlo. The bias for R5 (0.29 cmH 2 O/L/s; LoA: -2.69, 3.26) was broad and had random errors. The biases for whole-breath X5 (0.8 cmH 2 O/L/s; 95% LoA: -1.25, 2.85) was also wide and AX (-13.66 cmH 2 O/L/s; LoA: -45.46, 18.14) exhibited systematic errors. Zrs measurement differed significantly between the two FOT devices, particularly for reactance parameters. Therefore, normative values derived from one device should be avoided when interpreting results from another. • Respiratory impedance obtained from two oscillometers shows wide random and systematic variability. • Normative equations are device-specific and should not be used to interpret results obtained from other devices. • Standardization across oscillometers is crucial to ensure robust and reliable interpretation

Pulmonary Function DiagnosticsVol. 1(1)
All India Institute of Medical Sciences Raipur (IN)
Openalex Percentile: Top 2%
Chronic Obstructive Pulmonary Disease (COPD) Research
36.44
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.