Design, Calibration, and On-Site Validation of an LCVR-Driven Fast-Tunable Lyot Filter for the YOGIS Coronagraph
The Lyot filter, a fundamental element of the Yunnan Observatories Coronagraph Green-line Imaging System (YOGIS) at Lijiang Observatory, utilizes a Liquid Crystal Variable Retarder (LCVR) for swift electrical modulation. This filter allows for precise observations of the coronal green line (Fe XIV, central wavelength 5303 Å) with a narrow full-width at half-maximum (FWHM) of 1 Å and enables rapid adjustment of the transmission band wavelength. This feature aids in capturing the sky background intensity around the green line and images of two line wings (offset by ±0.45 Å from the central wavelength), crucial for determining the green line’s Doppler shift. By employing sky background subtraction and processing line wing images, an improved signal-to-noise ratio (SNR) in coronal green line images is achieved. The YOGIS Lyot filter, an enhancement of the NOrikura Green-line Imaging System (NOGIS) filter, operates at a wavelength of 5303 Å, offers a wavelength tuning range of ±2 Å, and tunes within <60 ms. This study elucidates the filter’s design principles, outlines essential calibration procedures, and validates its performance through on-site observations using the YOGIS.
Authors
- S. S. Su (ORCID: https://orcid.org/0000-0002-6569-3923)
- Z. Y. Li
- Yu Liu (ORCID: https://orcid.org/0000-0002-1869-5811)
- Mingyu Zhao
- Xuefei Zhang
Institutions
- Chinese Academy of Sciences (CN)
- Yunnan Observatories (CN)
- Southwest Jiaotong University (CN)
- Nanjing University (CN)
Publication Details
- Journal
- Photonics
- Published
- 2026-01-16
- DOI
- https://doi.org/10.3390/photonics13010076
- Primary Topic
- Solar and Space Plasma Dynamics
- Type
- article
- Field-Weighted Citation Impact
- 0.00