Laser, irrigation, and suction settings to optimize intra-renal temperature and pressure during automated laser lithotripsy in an ex vivo study part I

This study evaluated the effects of various laser, irrigation, and suction parameters on intra-renal temperature (IRT) and pressure (IRP) using an ex vivo porcine kidney. We introduced a novel Laser: Stop ratio (LS ratio), which automatically balances activation and resting periods. Only kidneys with a 10 mL pelvo-calyceal volume were used to assess the parameters [Thulium fiber laser (TFL) and Holmium: Yttrium-Aluminum-Garnet (Ho: YAG): 1.0 J/20Hz, 0.5 J/40Hz], pump-based irrigation of 30 mmHg (P30) and 50 mmHg (P50), LS ratios from 2:4 to 12:4, and continuous laser activation, with or without the flexible and navigable suction ureteral access sheath (FANS-UAS). With FANS-UAS, a Laser: Active suction: Rest ratio of 4:2:4 was applied at 200 mmHg suction. IRT and IRP were recorded every 30 s for 15 min, with dangerous thresholds set at > 50 °C for over 1 min or > 56 °C at any time. Under P30 and P50, the IRP remained below safety thresholds. Regarding IRT, for P30, TFL (1.0 J/20Hz) allowed unlimited pop-dusting at lower LS ratios (2:4), while TFL (0.5 J/40Hz) supported this even at higher LS ratios (4:4). Ho: YAG enabled longer pop-dusting durations, particularly at higher LS ratios (6:4) with adequate resting time. For P50, pop-dusting was unlimited at all settings with an LS ratio of 12:4 with adequate resting time. FANS-UAS allowed continuous pop-dusting. The LS ratio may offer insights into automatic laser fragmentation by balancing safety and efficiency. These findings suggest practical strategies for automated laser activation in robot-assisted flexible ureteroscopy.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2026-01-14
DOI
https://doi.org/10.1038/s41598-025-25265-0
Primary Topic
Kidney Stones and Urolithiasis Treatments
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Laser, irrigation, and suction settings to optimize intra-renal temperature and pressure during automated laser lithotripsy in an ex vivo study part I

Sang Won So, Christine Joy Castillo, Omar A Sulaiman, Fahad Alzahrani et al.
Scientific Reports
Kidney Stones and Urolithiasis Treatments
article

Laser, irrigation, and suction settings to optimize intra-renal temperature and pressure during automated laser lithotripsy in an ex vivo study part I

Sang Won So, Christine Joy Castillo, Omar A Sulaiman, Fahad Alzahrani, Dyandra Parikesit, Mohammed Almujhim, H.J. Lee, Majed Alharthi, Nasser Aljameel, Mohammad Zogan, Paul Renan Burgos Villegas, Ching-Heung Yen, Jose Carlo R. Elises, Jie-Hyeon Lee, Gyu Min Lee, Sung Yong Cho
article en

Abstract

This study evaluated the effects of various laser, irrigation, and suction parameters on intra-renal temperature (IRT) and pressure (IRP) using an ex vivo porcine kidney. We introduced a novel Laser: Stop ratio (LS ratio), which automatically balances activation and resting periods. Only kidneys with a 10 mL pelvo-calyceal volume were used to assess the parameters [Thulium fiber laser (TFL) and Holmium: Yttrium-Aluminum-Garnet (Ho: YAG): 1.0 J/20Hz, 0.5 J/40Hz], pump-based irrigation of 30 mmHg (P30) and 50 mmHg (P50), LS ratios from 2:4 to 12:4, and continuous laser activation, with or without the flexible and navigable suction ureteral access sheath (FANS-UAS). With FANS-UAS, a Laser: Active suction: Rest ratio of 4:2:4 was applied at 200 mmHg suction. IRT and IRP were recorded every 30 s for 15 min, with dangerous thresholds set at > 50 °C for over 1 min or > 56 °C at any time. Under P30 and P50, the IRP remained below safety thresholds. Regarding IRT, for P30, TFL (1.0 J/20Hz) allowed unlimited pop-dusting at lower LS ratios (2:4), while TFL (0.5 J/40Hz) supported this even at higher LS ratios (4:4). Ho: YAG enabled longer pop-dusting durations, particularly at higher LS ratios (6:4) with adequate resting time. For P50, pop-dusting was unlimited at all settings with an LS ratio of 12:4 with adequate resting time. FANS-UAS allowed continuous pop-dusting. The LS ratio may offer insights into automatic laser fragmentation by balancing safety and efficiency. These findings suggest practical strategies for automated laser activation in robot-assisted flexible ureteroscopy.

Scientific ReportsVol. 16(1)
Philippine General Hospital (PH), Seoul National University Hospital (KR), University of Indonesia (ID), Prince Mohammed bin Abdulaziz Hospital (SA), José R. Reyes Memorial Medical Center (PH), King Fahad Hospital Jeddah (SA), Tri-Service General Hospital Songshan Branch (TW), King Faisal University (SA), University of San Carlos (PH)
Korea Health Industry Development Institute
Openalex Percentile: Top 89%
Kidney Stones and Urolithiasis Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.