Feasibility of 3D-Printed Brachytherapy Contrast Markers for Use With CT Planning of Interstitial Procedures

Background Interstitial brachytherapy needles utilize radio-opaque markers that are expensive, not MR-safe, and have significant artifacts on CT/MRI. Here, we demonstrate the feasibility of 3D-printed radio-opaque markers composed of metal-infused polylactic acid (PLA) and test their use on CT-based localization of interstitial brachytherapy needles. Methodology Radio-opaque markers were 3D-printed using PLA filament infused with copper. Solid and hollow markers were designed, with hollow markers capable of containing viscous contrast media. Hollow markers were constructed heterogeneously with a single stripe of copper PLA along the length of a metal-free PLA to allow for feasibility testing of detecting dual contrast media using both CT and MRI. Markers were imaged via CT inside a prostate phantom and compared directly against steel and nitinol markers. Comparisons were performed in both soft tissue and air density regions. CT metal artifact (CTMA) was computed via standard deviation in the volume around each marker. Contrast-to-noise ratios (CNRs) were evaluated using CT imaging for all tested markers using identical methodology. Artifact differences between markers were compared by evaluating the standard deviations of fixed volumes around each voxel. Results The average CNR for solid and hollow copper PLA markers was 1.99 ± 0.07 and 1.23 ± 0.21 in tissue and 2.48 ± 0.10 and 2.99 ± 0.15 in air, respectively. The average CNR for wire and nitinol markers was 1.31 ± 0.02 and 2.33 ± 0.05 in tissue and 1.86 ± 0.01 and 2.76 ± 0.01 in air, respectively. CTMA for solid and hollow copper PLA markers were 13.62 and 8.80 HU in tissue and 9.99 and 13.66 HU in air, respectively. The average artifact for the nitinol and wire markers was 75.93 and 18.06 HU in tissue and 50.71 and 16.64 HU in air, respectively. Conclusions Solid copper PLA 3D-printed markers have superior visibility to commercially available steel marker wires. Both types of printed markers exhibited reduced imaging artifacts when compared with commercially available markers. Printed markers can improve the accuracy of interstitial planning and can be inexpensively produced in-house.

Authors

Institutions

Publication Details

Journal
Cureus
Published
2026-01-14
DOI
https://doi.org/10.7759/cureus.101524
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Feasibility of 3D-Printed Brachytherapy Contrast Markers for Use With CT Planning of Interstitial Procedures

Bart Frizzell, Anna C. Snavely, Travis G. Marshall, Megan E. Lipford et al.
Cureus
Advanced X-ray and CT Imaging
article

Feasibility of 3D-Printed Brachytherapy Contrast Markers for Use With CT Planning of Interstitial Procedures

Bart Frizzell, Anna C. Snavely, Travis G. Marshall, Megan E. Lipford, James D. Ververs, N. Razavian, M McKee, Ryan T. Hughes, W. Dolesh, S.E. Glynn, India R Wood, Cody Lackey, Doris R Brown, Paul J Black, Michael Farris
article en

Abstract

Background Interstitial brachytherapy needles utilize radio-opaque markers that are expensive, not MR-safe, and have significant artifacts on CT/MRI. Here, we demonstrate the feasibility of 3D-printed radio-opaque markers composed of metal-infused polylactic acid (PLA) and test their use on CT-based localization of interstitial brachytherapy needles. Methodology Radio-opaque markers were 3D-printed using PLA filament infused with copper. Solid and hollow markers were designed, with hollow markers capable of containing viscous contrast media. Hollow markers were constructed heterogeneously with a single stripe of copper PLA along the length of a metal-free PLA to allow for feasibility testing of detecting dual contrast media using both CT and MRI. Markers were imaged via CT inside a prostate phantom and compared directly against steel and nitinol markers. Comparisons were performed in both soft tissue and air density regions. CT metal artifact (CTMA) was computed via standard deviation in the volume around each marker. Contrast-to-noise ratios (CNRs) were evaluated using CT imaging for all tested markers using identical methodology. Artifact differences between markers were compared by evaluating the standard deviations of fixed volumes around each voxel. Results The average CNR for solid and hollow copper PLA markers was 1.99 ± 0.07 and 1.23 ± 0.21 in tissue and 2.48 ± 0.10 and 2.99 ± 0.15 in air, respectively. The average CNR for wire and nitinol markers was 1.31 ± 0.02 and 2.33 ± 0.05 in tissue and 1.86 ± 0.01 and 2.76 ± 0.01 in air, respectively. CTMA for solid and hollow copper PLA markers were 13.62 and 8.80 HU in tissue and 9.99 and 13.66 HU in air, respectively. The average artifact for the nitinol and wire markers was 75.93 and 18.06 HU in tissue and 50.71 and 16.64 HU in air, respectively. Conclusions Solid copper PLA 3D-printed markers have superior visibility to commercially available steel marker wires. Both types of printed markers exhibited reduced imaging artifacts when compared with commercially available markers. Printed markers can improve the accuracy of interstitial planning and can be inexpensively produced in-house.

CureusVol. 18(1)
Atrium Health Wake Forest Baptist (US), Novant Health (US), Moffitt Cancer Center (US)
Sustainable cities and communities
Openalex Percentile: Top 88%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.