Robustness of LET <sub>d</sub> -optimized multi-ion therapy against range and setup uncertainties: evaluation and enhancement with carbon-, oxygen-, and neon-ion beams

<i>Objective.</i>The LET trilemma-an inherent conflict between target dose homogeneity, range robustness, and high dose-averaged linear energy transfer (LET<sub>d</sub>)-poses a major challenge in treatment optimization. To ensure accurate beam delivery in multi-ion therapy, this study evaluated the effects of range and setup uncertainties on LET<sub>d</sub>-optimized treatment plans and explored strategies to overcome this trilemma, framed within the phase I LET<sub>d</sub>escalation trial for head and neck cancers.<i>Approach.</i>Six head and neck cancer patients representing diverse tumors were selected. Multi-ion therapy plans using carbon-, oxygen-, and neon-ion beams were optimized to achieve a target LET<sub>d</sub>of 90 keV μm<sup>-1</sup>(the final LET<sub>d</sub>level of the phase I trial). These plans were recalculated to incorporate systematic range uncertainty (±2.5%) and random daily setup variations (mean, 0.45 mm; standard deviation, 0.23 mm) across the 16 fractions, and their combined effects on the dose and LET<sub>d</sub>distributions were evaluated. Additionally, to explore strategies to enhance plan robustness, five modified plans were evaluated for one patient identified as particularly susceptible to these uncertainties.<i>Main Results.</i>Range uncertainty was the dominant contributor to degraded plan quality, substantially outweighing setup uncertainty. A small, centrally located tumor was most susceptible, exhibiting dose inhomogeneity of approximately 11%, while LET<sub>d</sub>variations were approximately 3 keV μm<sup>-1</sup>. The most effective mitigation strategy involved replacing the original carbon-oxygen combination with oxygen ions for two beam ports, reducing dose inhomogeneity by more than 7% while maintaining normal tissue sparing adjacent to the target.<i>Significance.</i>Optimization toward achieving higher LET<sub>d</sub>makes treatment plans susceptible to range uncertainty, leading to dose degradation within small, deep-seated tumors. Employing heavier ions is an effective strategy to overcome this challenge, enabling robust target coverage by leveraging their inherently higher LET<sub>d</sub>while sparing normal tissues. These findings provide a key rationale for ion selection in the design of robust multi-ion therapy.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2026-01-14
DOI
https://doi.org/10.1088/1361-6560/ae387b
Primary Topic
Radiation Therapy and Dosimetry
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Robustness of LET d -optimized multi-ion therapy against range and setup uncertainties: evaluation and enhancement with carbon-, oxygen-, and neon-ion beams

Yusuke Nomura, Takamitsu Masuda, Koki Kasamatsu, Masashi Koto et al.
Physics in Medicine and Biology
Radiation Therapy and Dosimetry
article

Robustness of LET d -optimized multi-ion therapy against range and setup uncertainties: evaluation and enhancement with carbon-, oxygen-, and neon-ion beams

Yusuke Nomura, Takamitsu Masuda, Koki Kasamatsu, Masashi Koto, H. Ikawa, Nobuyuki Kanematsu, Makoto Shinoto, Taku Inaniwa
article en

Abstract

<i>Objective.</i>The LET trilemma-an inherent conflict between target dose homogeneity, range robustness, and high dose-averaged linear energy transfer (LET<sub>d</sub>)-poses a major challenge in treatment optimization. To ensure accurate beam delivery in multi-ion therapy, this study evaluated the effects of range and setup uncertainties on LET<sub>d</sub>-optimized treatment plans and explored strategies to overcome this trilemma, framed within the phase I LET<sub>d</sub>escalation trial for head and neck cancers.<i>Approach.</i>Six head and neck cancer patients representing diverse tumors were selected. Multi-ion therapy plans using carbon-, oxygen-, and neon-ion beams were optimized to achieve a target LET<sub>d</sub>of 90 keV μm<sup>-1</sup>(the final LET<sub>d</sub>level of the phase I trial). These plans were recalculated to incorporate systematic range uncertainty (±2.5%) and random daily setup variations (mean, 0.45 mm; standard deviation, 0.23 mm) across the 16 fractions, and their combined effects on the dose and LET<sub>d</sub>distributions were evaluated. Additionally, to explore strategies to enhance plan robustness, five modified plans were evaluated for one patient identified as particularly susceptible to these uncertainties.<i>Main Results.</i>Range uncertainty was the dominant contributor to degraded plan quality, substantially outweighing setup uncertainty. A small, centrally located tumor was most susceptible, exhibiting dose inhomogeneity of approximately 11%, while LET<sub>d</sub>variations were approximately 3 keV μm<sup>-1</sup>. The most effective mitigation strategy involved replacing the original carbon-oxygen combination with oxygen ions for two beam ports, reducing dose inhomogeneity by more than 7% while maintaining normal tissue sparing adjacent to the target.<i>Significance.</i>Optimization toward achieving higher LET<sub>d</sub>makes treatment plans susceptible to range uncertainty, leading to dose degradation within small, deep-seated tumors. Employing heavier ions is an effective strategy to overcome this challenge, enabling robust target coverage by leveraging their inherently higher LET<sub>d</sub>while sparing normal tissues. These findings provide a key rationale for ion selection in the design of robust multi-ion therapy.

Physics in Medicine and BiologyVol. 71(3)
Yamagata University (JP), National Institutes for Quantum Science and Technology (JP)
Affordable and clean energy
Openalex Percentile: Top 89%
Radiation Therapy and Dosimetry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.