Sharpening the surgeon’s eye: an adaptable dual-mode gamma probe architecture optimized for high-resolution and high-sensitivity radio-guided surgery

<i>Objective.</i>To design and validate a single, reconfigurable gamma probe that overcomes the static compromise between spatial resolution and sensitivity in radio-guided surgery, enabling both rapid lesion detection and precise margin delineation.<i>Approach.</i>A dual-layer lead collimator was designed for a LaBr₃(Ce)-SiPM detector. A validated analytical model coupled with a multi-objective genetic algorithm (NSGA-II) was used to explore the theoretical performance limits and identify optimal geometries. A two-phase computational search identified a single, universal geometry that can be switched intraoperatively between a high-sensitivity (HS) mode and a high-resolution (HR) mode by adjusting collimator positions.<i>Main results.</i>The universal design, at a 30 mm distance, achieves a spatial resolution of 6.41 mm full width at half maximum (FWHM) in HR mode and a sensitivity of 1483 counts per second (cps)/MBq in HS mode. The optimization framework identified specialized, distance-specific theoretical designs with resolutions as fine as 3.26 mm FWHM. The underlying detector's energy resolution is sufficient to distinguish between ⁹⁹<sub>m</sub>Tc (140.5 keV) and<sup>123</sup>I (159 keV).<i>Significance.</i>This work presents a practical, single-instrument solution that offers surgeons the intraoperative flexibility to prioritize either rapid detection or precise delineation. The developed design methodology provides a robust framework for creating next-generation, application-specific surgical guidance tools.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2026-01-14
DOI
https://doi.org/10.1088/1361-6560/ae387e
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Sharpening the surgeon’s eye: an adaptable dual-mode gamma probe architecture optimized for high-resolution and high-sensitivity radio-guided surgery

Ahmet Yılmaz, Muhammed Emin Bedir, Bruce R Thomadsen
Physics in Medicine and Biology
Advanced Radiotherapy Techniques
article

Sharpening the surgeon’s eye: an adaptable dual-mode gamma probe architecture optimized for high-resolution and high-sensitivity radio-guided surgery

Ahmet Yılmaz, Muhammed Emin Bedir, Bruce R Thomadsen
article en

Abstract

<i>Objective.</i>To design and validate a single, reconfigurable gamma probe that overcomes the static compromise between spatial resolution and sensitivity in radio-guided surgery, enabling both rapid lesion detection and precise margin delineation.<i>Approach.</i>A dual-layer lead collimator was designed for a LaBr₃(Ce)-SiPM detector. A validated analytical model coupled with a multi-objective genetic algorithm (NSGA-II) was used to explore the theoretical performance limits and identify optimal geometries. A two-phase computational search identified a single, universal geometry that can be switched intraoperatively between a high-sensitivity (HS) mode and a high-resolution (HR) mode by adjusting collimator positions.<i>Main results.</i>The universal design, at a 30 mm distance, achieves a spatial resolution of 6.41 mm full width at half maximum (FWHM) in HR mode and a sensitivity of 1483 counts per second (cps)/MBq in HS mode. The optimization framework identified specialized, distance-specific theoretical designs with resolutions as fine as 3.26 mm FWHM. The underlying detector's energy resolution is sufficient to distinguish between ⁹⁹<sub>m</sub>Tc (140.5 keV) and<sup>123</sup>I (159 keV).<i>Significance.</i>This work presents a practical, single-instrument solution that offers surgeons the intraoperative flexibility to prioritize either rapid detection or precise delineation. The developed design methodology provides a robust framework for creating next-generation, application-specific surgical guidance tools.

Physics in Medicine and BiologyVol. 71(3)
University of Wisconsin–Madison (US), Karamanoğlu Mehmetbey University (TR)
Openalex Percentile: Top 88%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.