Durable low-friction plasma-polymerized coating for suture needles ensuring safety

• One-step atmospheric-pressure plasma polymerization of TRIAP on curved suture needles. • A 5-minute plasma treatment reduces needle penetration forces by up to 35%. • Coating remains intact over multiple insertions, maintaining low friction. • XPS reveals increased SiO 4 and NH 2 bonds, reducing friction and improving adhesion. • No new volatile compounds are released from coated needles, confirming safety. In wound suturing, the needle and thread are crucial. The ideal surgical needle must be rigid yet flexible, corrosion-resistant, sterile and effectively lubricated to penetrate tissue effortlessly and minimize damage. However, achieving optimal lubrication with current methods, involving cumbersome silicone application and high-temperature curing (2–6 h at 150–250 °C), remains challenging due to the lengthy process. This study explored the use of an atmospheric-pressure plasma jet (APPJ) to rapidly apply a stable, low-friction, one-step coating on curved suture needles. Plasma-polymerized coatings based on N 1 -(3-trimethoxysilylpropyl)diethylenetriamine (TRIAP) were applied. The influence of plasma exposure time was analyzed. After 5 min of plasma treatment, coated needles showed up to a 35 % reduction in penetration forces compared to control needles (silicone-coated). Longer plasma exposure times enhanced the coating’s low-friction and adhesion properties by promoting the formation of SiO 4 and NH 2 bonds, respectively. A minimum treatment time ensured an even coating across the needle surface. Notably, the plasma-polymerized coatings did not introduce any new volatile or semi-volatile compounds, maintaining the needle’s chemical profile. This study demonstrates a faster, more efficient one-step coating process that is scalable to industry, offering consistent performance while significantly reducing suturing forces compared to the conventional silicone coating.

Authors

Institutions

Publication Details

Journal
Applied Surface Science
Published
2026-01-14
DOI
https://doi.org/10.1016/j.apsusc.2026.165930
Primary Topic
Surgical Sutures and Adhesives
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Durable low-friction plasma-polymerized coating for suture needles ensuring safety

Marina Corral-Bobadilla, Fernando Alba‐Elías, Elisa Sainz‐García, Ana Rodríguez Bernaldo de Quirós et al.
Applied Surface Science
Surgical Sutures and Adhesives
article

Durable low-friction plasma-polymerized coating for suture needles ensuring safety

Marina Corral-Bobadilla, Fernando Alba‐Elías, Elisa Sainz‐García, Ana Rodríguez Bernaldo de Quirós, Rodolfo Múgica‐Vidal, Patricia Vázquez Loureiro, Letricia Barbosa‐Pereira, Ignacio Muro-Fraguas, Antía Lestido-Cardama, Miguel Moreno-Martínez, Ana Sainz-García
article en

Abstract

• One-step atmospheric-pressure plasma polymerization of TRIAP on curved suture needles. • A 5-minute plasma treatment reduces needle penetration forces by up to 35%. • Coating remains intact over multiple insertions, maintaining low friction. • XPS reveals increased SiO 4 and NH 2 bonds, reducing friction and improving adhesion. • No new volatile compounds are released from coated needles, confirming safety. In wound suturing, the needle and thread are crucial. The ideal surgical needle must be rigid yet flexible, corrosion-resistant, sterile and effectively lubricated to penetrate tissue effortlessly and minimize damage. However, achieving optimal lubrication with current methods, involving cumbersome silicone application and high-temperature curing (2–6 h at 150–250 °C), remains challenging due to the lengthy process. This study explored the use of an atmospheric-pressure plasma jet (APPJ) to rapidly apply a stable, low-friction, one-step coating on curved suture needles. Plasma-polymerized coatings based on N 1 -(3-trimethoxysilylpropyl)diethylenetriamine (TRIAP) were applied. The influence of plasma exposure time was analyzed. After 5 min of plasma treatment, coated needles showed up to a 35 % reduction in penetration forces compared to control needles (silicone-coated). Longer plasma exposure times enhanced the coating’s low-friction and adhesion properties by promoting the formation of SiO 4 and NH 2 bonds, respectively. A minimum treatment time ensured an even coating across the needle surface. Notably, the plasma-polymerized coatings did not introduce any new volatile or semi-volatile compounds, maintaining the needle’s chemical profile. This study demonstrates a faster, more efficient one-step coating process that is scalable to industry, offering consistent performance while significantly reducing suturing forces compared to the conventional silicone coating.

Applied Surface ScienceVol. 726
Universidad de La Rioja (ES), Universidad Pública de Navarra (UPNA) (ES), Universidade de Santiago de Compostela (ES)
Industry, innovation and infrastructure
Openalex Percentile: Top 89%
Surgical Sutures and Adhesives
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.