'Schanz Pin Extractor' Technique for Removal of a Broken Intramedullary Nail
Abstract Intramedullary nail removal is an uncommon but complicated procedure with significant surgical challenges, especially in the case of a broken nail. There are multiple cumbersome techniques for nail removal described that are frought with difficulty. We present a technique designed to take advantage of the difference in metallurgy of a stainless steel Schanz pin and a cannulated titanium intramedullary nail. We use a 5.0 mm self-tapping Schanz pin that is able to be drilled into the softer cannulated segment of titanium intramedullary nail designed to accomate a guide wire. Once drilled into the desired segment of nail, the distal interlock can be removed, and the Schanz pin can be connected to a T-shaped handle used as an extractor to pull out the nail.
Authors
- Emmanuele Santolini (ORCID: https://orcid.org/0000-0002-0899-8317)
- Daniel Scott Horwitz (ORCID: https://orcid.org/0000-0001-5139-3308)
- Lamberto Felli
- Austin Kloc
Institutions
- Geisinger Medical Center (US)
- Ospedale Policlinico San Martino (IT)
- University of Genoa (IT)
Publication Details
- Journal
- Zenodo (CERN European Organization for Nuclear Research)
- Published
- 2025-12-23
- DOI
- https://doi.org/10.5281/zenodo.18211399
- Primary Topic
- Nail Diseases and Treatments
- Type
- article
- Field-Weighted Citation Impact
- 0.00