Biomechanical evaluation of reconstruction plates in bone-grafted mandibular defects: a finite element analysis
Thicker reconstruction plates enhance mechanical resistance but increase stress concentration on fixation screws, especially near osteotomies, suggesting a higher risk of screw loosening or peri‑implant bone resorption. The anterior mandible demonstrated the most unfavorable stress distribution. Locking fixation systems are recommended when rigid plates are used to mitigate screw-related complications.