Heat generation of autologous bone harvesting drills: an in vitro study

Bone collecting drills are utilized in implantology to harvest autologous bone chips. This study aimed to assess intraosseous and bone chip temperature changes during bone collecting procedures. Auto-Max (Megagen, Luton, UK) drills were evaluated at rotational speeds of 300, 600, 1200, and 2000 revolutions per minute (rpm), each combined with axial loads of 15 N, 20 N, and 25 N, using fresh pig ribs. Axial load along the perpendicular axis was maintained via a dedicated drilling tower. Bone chip temperatures were measured using an infrared non-contact thermometer, whereas donor bone temperatures were monitored with thermocouple sensors positioned 0.5 mm from the osteotomy periphery. Drill wear was assessed through scanning electron microscopy. The highest donor bone temperature observed was 10.1 ± 3.04 °C at 1200 rpm and 25 N. With a 20 N load, temperature increases ranged from 3.01 to 4.84 °C. At 25 N, temperatures at 300 rpm (1.84 ± 0.58 °C) and 600 rpm (6.70 ± 2.64 °C) consistently stayed below 10 °C. Bone chip temperature rises were all under 5 °C. The slowest drilling occurred at 15 N/600 rpm (6.49s), and the fastest at 2000 rpm (< 1.65s). Drill wear was moderate up to 10 uses, but significant after 30, with bone temperatures rising by 240% to an average of 10.22 ± 2.45 °C. With respect to chip temperatures, all evaluated drilling parameters produced maximum temperature elevations below 5 °C. Donor bone temperature increases remained consistently under the theoretical threshold of 10 °C when an axial load of 20 N was applied-regardless of rotational speed-or when drilling at 600 rpm, independent of axial load. It is recommended that drills be replaced prior to 30 uses.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2026-01-13
DOI
https://doi.org/10.1038/s41598-026-35988-3
Primary Topic
Dental Implant Techniques and Outcomes
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Heat generation of autologous bone harvesting drills: an in vitro study

Stephan Christian Möhlhenrich, Edina Lempel, József Szalma, Nick Köhler et al.
Scientific Reports
Dental Implant Techniques and Outcomes
article

Heat generation of autologous bone harvesting drills: an in vitro study

Stephan Christian Möhlhenrich, Edina Lempel, József Szalma, Nick Köhler, Fanni Jáni, Péter Szabó
article en

Abstract

Bone collecting drills are utilized in implantology to harvest autologous bone chips. This study aimed to assess intraosseous and bone chip temperature changes during bone collecting procedures. Auto-Max (Megagen, Luton, UK) drills were evaluated at rotational speeds of 300, 600, 1200, and 2000 revolutions per minute (rpm), each combined with axial loads of 15 N, 20 N, and 25 N, using fresh pig ribs. Axial load along the perpendicular axis was maintained via a dedicated drilling tower. Bone chip temperatures were measured using an infrared non-contact thermometer, whereas donor bone temperatures were monitored with thermocouple sensors positioned 0.5 mm from the osteotomy periphery. Drill wear was assessed through scanning electron microscopy. The highest donor bone temperature observed was 10.1 ± 3.04 °C at 1200 rpm and 25 N. With a 20 N load, temperature increases ranged from 3.01 to 4.84 °C. At 25 N, temperatures at 300 rpm (1.84 ± 0.58 °C) and 600 rpm (6.70 ± 2.64 °C) consistently stayed below 10 °C. Bone chip temperature rises were all under 5 °C. The slowest drilling occurred at 15 N/600 rpm (6.49s), and the fastest at 2000 rpm (< 1.65s). Drill wear was moderate up to 10 uses, but significant after 30, with bone temperatures rising by 240% to an average of 10.22 ± 2.45 °C. With respect to chip temperatures, all evaluated drilling parameters produced maximum temperature elevations below 5 °C. Donor bone temperature increases remained consistently under the theoretical threshold of 10 °C when an axial load of 20 N was applied-regardless of rotational speed-or when drilling at 600 rpm, independent of axial load. It is recommended that drills be replaced prior to 30 uses.

Scientific ReportsVol. 16(1)
Radboud University Nijmegen (NL), Radboud University Medical Center (NL), Szent János Kórház (HU), University of Pecs (HU)
Pécsi Tudományegyetem, Általános Orvostudományi Kar, Pécsi Tudományegyetem
Openalex Percentile: Top 89%
Dental Implant Techniques and Outcomes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.