Damage to Polymer Coatings on Microguidewire Tips through Shaping and J-wiring for Optimizing Flow Diverter Deployment: A Scanning Electron Microscopy Study

Achieving adequate wall apposition is a crucial technical goal when deploying flow diverters to treat wide-neck cerebral aneurysms. The socalled J-wiring technique is a common method used to optimize flow diverter wall apposition. However, the frictional interaction between the shaping device and the microguidewire tip during the formation of the J-loop, as well as the interaction between the J-loop and the flow diverter during J-wiring, may potentially cause damage to the guidewire's coating. Three frequently used guidewires were tested in vitro in a silicone aneurysm model. Manual J-shaping of guidewire tips, along with the J-wiring technique (including J-shaping), caused damage to the surface coating of guidewires, as observed by scanning electron microscopy. Therefore, both mechanisms may contribute to the generation of polymer micro-fragments in patients treated with flow diversion.

Authors

Institutions

Publication Details

Journal
Neurointervention
Published
2026-01-12
DOI
https://doi.org/10.5469/neuroint.2025.01130
Primary Topic
Cerebrospinal fluid and hydrocephalus
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Damage to Polymer Coatings on Microguidewire Tips through Shaping and J-wiring for Optimizing Flow Diverter Deployment: A Scanning Electron Microscopy Study

Esben Thormann, Rasmus Holmboe Dahl, René Wugt Larsen, Goetz Benndorf
Neurointervention
Cerebrospinal fluid and hydrocephalus
article

Damage to Polymer Coatings on Microguidewire Tips through Shaping and J-wiring for Optimizing Flow Diverter Deployment: A Scanning Electron Microscopy Study

Esben Thormann, Rasmus Holmboe Dahl, René Wugt Larsen, Goetz Benndorf
article en

Abstract

Achieving adequate wall apposition is a crucial technical goal when deploying flow diverters to treat wide-neck cerebral aneurysms. The socalled J-wiring technique is a common method used to optimize flow diverter wall apposition. However, the frictional interaction between the shaping device and the microguidewire tip during the formation of the J-loop, as well as the interaction between the J-loop and the flow diverter during J-wiring, may potentially cause damage to the guidewire's coating. Three frequently used guidewires were tested in vitro in a silicone aneurysm model. Manual J-shaping of guidewire tips, along with the J-wiring technique (including J-shaping), caused damage to the surface coating of guidewires, as observed by scanning electron microscopy. Therefore, both mechanisms may contribute to the generation of polymer micro-fragments in patients treated with flow diversion.

NeurointerventionVol. 21(1)
Baylor College of Medicine (US), Copenhagen University Hospital (DK), Rigshospitalet (DK), Technical University of Denmark (DK)
Faculty of Health and Medical Sciences, University of Western Australia, Det Sundhedsvidenskabelige Fakultet, Københavns Universitet
Openalex Percentile: Top 91%
Cerebrospinal fluid and hydrocephalus
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.