Multi-wavelength optical sensor for jaundice detection via absorbance difference spectroscopy

We demonstrate a low-cost optical sensor for neonatal jaundice detection using 470 nm (bilirubin-sensitive) and 560 nm (hemoglobin isosbestic) LEDs. The system applies absorbance difference spectroscopy to isolate bilirubin signals in skin-mimicking phantoms, achieving 387<i>m</i><i>V</i>/<i>m</i><i>g</i>⋅<i>d</i><i>L</i><sup>-1</sup> sensitivity and 13.5% error versus spectrophotometry. Optical characterization confirms 470 nm penetration depth (≈0.65<i>m</i><i>m</i>) in neonatal dermis. A machine learning classifier predicts risk severity with 99.5% accuracy. Comparative analysis shows a superior cost-performance trade-off versus reflectance-based devices. This work enables accessible hyperbilirubinemia monitoring through optimized wavelength selection and transmission-mode detection.

Authors

Institutions

Publication Details

Journal
Applied Optics
Published
2026-01-12
DOI
https://doi.org/10.1364/ao.575790
Primary Topic
Neonatal Health and Biochemistry
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Multi-wavelength optical sensor for jaundice detection via absorbance difference spectroscopy

SARA ALSAFRAN, Gameel Saleh, DOAA BAHBRY, MARWA ALOTAIBY et al.
Applied Optics
Neonatal Health and Biochemistry
article

Multi-wavelength optical sensor for jaundice detection via absorbance difference spectroscopy

SARA ALSAFRAN, Gameel Saleh, DOAA BAHBRY, MARWA ALOTAIBY, WALA’A ALMURAIKHI
article en

Abstract

We demonstrate a low-cost optical sensor for neonatal jaundice detection using 470 nm (bilirubin-sensitive) and 560 nm (hemoglobin isosbestic) LEDs. The system applies absorbance difference spectroscopy to isolate bilirubin signals in skin-mimicking phantoms, achieving 387<i>m</i><i>V</i>/<i>m</i><i>g</i>⋅<i>d</i><i>L</i><sup>-1</sup> sensitivity and 13.5% error versus spectrophotometry. Optical characterization confirms 470 nm penetration depth (≈0.65<i>m</i><i>m</i>) in neonatal dermis. A machine learning classifier predicts risk severity with 99.5% accuracy. Comparative analysis shows a superior cost-performance trade-off versus reflectance-based devices. This work enables accessible hyperbilirubinemia monitoring through optimized wavelength selection and transmission-mode detection.

Applied OpticsVol. 65(4)
Taibah University (SA), Imam Abdulrahman Bin Faisal University (SA)
Openalex Percentile: Top 91%
Neonatal Health and Biochemistry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.