Automated Thromboelastometry for Semiquantitative Assessment of High Heparin Concentrations: A Feasibility Study in Healthy Volunteers
This study on automated thromboelastometry demonstrated a dose-dependent prolongation of HEPTEM CT when heparin concentration increased from 0 to 6 IU/mL. Furthermore, ROC analysis on the ability of HEPTEM CT to detect heparin concentrations of ≥ 4 IU/mL indicated possible diagnostic potential. These novel findings suggest that HEPTEM CT is a feasible ROTEM Sigma parameter for semiquantitative assessment of heparin concentrations required for safe CPB. However, this remains to be verified in future clinical studies.
Authors
- Timo Kaakinen (ORCID: https://orcid.org/0000-0001-5125-245X)
- Tiina Erkinaro (ORCID: https://orcid.org/0000-0002-4566-1887)
- Eeva‐Riitta Savolainen (ORCID: https://orcid.org/0000-0002-7666-4597)
- Pasi Ohtonen (ORCID: https://orcid.org/0000-0001-8245-3000)
- Marika Pada
- Janne Moilanen
- Janne Liisanantti
Institutions
- Oulu University Hospital (FI)
- Vaasa Central Hospital (FI)
Publication Details
- Journal
- Health Science Reports
- Published
- 2026-01-01
- DOI
- https://doi.org/10.1002/hsr2.71721
- Primary Topic
- Trauma, Hemostasis, Coagulopathy, Resuscitation
- Type
- article
- Field-Weighted Citation Impact
- 0.00