An Artificial Intelligence‐Based Computer Vision Model for Human Sperm Concentration, Motility, and Kinematics Analysis

Accurate assessment of sperm concentration and motility is critical for the diagnosis and management of male infertility. However, current methods, manual hemocytometer counting and commercial computer-aided sperm analysis (CASA) systems, are limited by labor intensity, human error, and variable performance under diverse sample conditions. Here, we present an artificial intelligence (AI)-driven computer vision tool for high-resolution, quantitative analysis of sperm motility and concentration. In a prospective study of 26 semen samples (22 patients, 4 donors), we benchmarked the AI model against manual tracking (using Fiji software) and a commercial CASA system (Hamilton Thorne IVOS II). Our method computed concentration and motility parameters, including straight-line velocity (VSL), curvilinear velocity (VCL), average path velocity (VAP), linearity (LIN), amplitude of lateral head displacement (ALH<sub>max</sub>), and beat cross frequency (BCF). Calibration using donor samples enabled accurate mapping of tracked sperm counts to concentrations. The AI tool presented a strong linear correlation with manual tracking (<i>R</i> <sup>2</sup> = 0.93-0.98; Root Mean Square Error (RMSE) = 3.3-7.3 μm/s for VSL, VCL, VAP), and outperformed CASA in both accuracy and consistency across all motility parameters. Post-calibration, ALH<sub>max</sub> and BCF estimates improved substantially, with a 30%-50% reduction in RMSE. Grading of sperm motility by the AI model aligned closely with manual classification, avoiding the systematic misclassification typically observed with CASA. Furthermore, the AI system exhibited higher repeatability and robustness across duplicate samples and variable imaging conditions, with deviations below ± 2%. These findings demonstrate that our AI-based tool offers a quantitative and reliable alternative to current semen analysis platforms, supporting improved fertility diagnostics and potentially a more informative treatment process.

Authors

Institutions

Publication Details

Journal
Smart Medicine
Published
2026-01-09
DOI
https://doi.org/10.1002/smmd.70026
Primary Topic
Sperm and Testicular Function
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

An Artificial Intelligence‐Based Computer Vision Model for Human Sperm Concentration, Motility, and Kinematics Analysis

Adrian Neild, Klaus Ackermann, Reza Nosrati, Sahar Shahali et al.
Smart Medicine
Sperm and Testicular Function
article

An Artificial Intelligence‐Based Computer Vision Model for Human Sperm Concentration, Motility, and Kinematics Analysis

Adrian Neild, Klaus Ackermann, Reza Nosrati, Sahar Shahali, RI McLachlan, Gulfam Ahmad, Deirdre Zander‐Fox, David Mortimer, Moira K. O'Bryan
article en

Abstract

Accurate assessment of sperm concentration and motility is critical for the diagnosis and management of male infertility. However, current methods, manual hemocytometer counting and commercial computer-aided sperm analysis (CASA) systems, are limited by labor intensity, human error, and variable performance under diverse sample conditions. Here, we present an artificial intelligence (AI)-driven computer vision tool for high-resolution, quantitative analysis of sperm motility and concentration. In a prospective study of 26 semen samples (22 patients, 4 donors), we benchmarked the AI model against manual tracking (using Fiji software) and a commercial CASA system (Hamilton Thorne IVOS II). Our method computed concentration and motility parameters, including straight-line velocity (VSL), curvilinear velocity (VCL), average path velocity (VAP), linearity (LIN), amplitude of lateral head displacement (ALH<sub>max</sub>), and beat cross frequency (BCF). Calibration using donor samples enabled accurate mapping of tracked sperm counts to concentrations. The AI tool presented a strong linear correlation with manual tracking (<i>R</i> <sup>2</sup> = 0.93-0.98; Root Mean Square Error (RMSE) = 3.3-7.3 μm/s for VSL, VCL, VAP), and outperformed CASA in both accuracy and consistency across all motility parameters. Post-calibration, ALH<sub>max</sub> and BCF estimates improved substantially, with a 30%-50% reduction in RMSE. Grading of sperm motility by the AI model aligned closely with manual classification, avoiding the systematic misclassification typically observed with CASA. Furthermore, the AI system exhibited higher repeatability and robustness across duplicate samples and variable imaging conditions, with deviations below ± 2%. These findings demonstrate that our AI-based tool offers a quantitative and reliable alternative to current semen analysis platforms, supporting improved fertility diagnostics and potentially a more informative treatment process.

Smart MedicineVol. 5(1)
Royal Children's Hospital (AU), The University of Melbourne (AU), Discovery Institute (US), Hudson Institute of Medical Research (AU), Monash IVF (AU), Monash University (AU), The University of Adelaide (AU)
Medical Research Council, National Health and Medical Research Council
Openalex Percentile: Top 93%
Sperm and Testicular Function
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.