Evaluation of the <scp>ForenSeq</scp> ™ <scp>MainstAY</scp> workflow for forensic <scp>DNA</scp> samples

This study evaluated the performance and limitations of the ForenSeq™ MainstAY chemistry on the MiSeq FGx® Sequencing System, which is approved for National DNA Index System (NDIS) use. This study included baseline performance, sensitivity and reproducibility studies, challenging casework samples, assessments of library plexities, comparisons with capillary electrophoresis (CE), standard versus enhanced PCR1 buffers, manual versus automated library preparation, and different flow cell types. To assess reproducibility across users, 33 samples were prepared by both an experienced and a novice user. Overall, the ForenSeq™ MainstAY Kit demonstrated high reliability with pristine samples and consistent performance across users, with minimal locus dropout until 62 pg. The kit outperformed CE for low-input and degraded samples, but limitations were observed at higher plexities. Increased sample numbers on a single flow cell disproportionately reduced read counts and locus recovery in degraded casework samples, while pristine samples were less affected. Strategic adjustments, such as batching degraded samples together or lowering plexity, may help mitigate these effects. Casework samples, including blood, touched items, bone, hair, and tissue, were tested with both standard and enhanced PCR1 buffers. In a 64-sample run (32 per buffer type), increased multiplexing reduced locus recovery in standard buffer samples, while the enhanced buffer improved recovery in 16 samples. Automated library preparation on the Opentrons OT-2 produced comparable or improved recovery relative to manual preparation, with no evidence of cross-contamination. These findings provide practical guidance for forensic laboratories adopting the MainstAY Kit, particularly in optimizing workflows for challenging samples and high-throughput sequencing environments.

Authors

Institutions

Publication Details

Journal
Journal of Forensic Sciences
Published
2026-01-09
DOI
https://doi.org/10.1111/1556-4029.70263
Primary Topic
Forensic and Genetic Research
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of the ForenSeq ™ MainstAY workflow for forensic DNA samples

Ryan Gutierrez, Rachel M. Houston, Lucio L. Avellaneda, Damani Johnson et al.
Journal of Forensic Sciences
Forensic and Genetic Research
article

Evaluation of the ForenSeq ™ MainstAY workflow for forensic DNA samples

Ryan Gutierrez, Rachel M. Houston, Lucio L. Avellaneda, Damani Johnson, Tim Kalafut
article en

Abstract

This study evaluated the performance and limitations of the ForenSeq™ MainstAY chemistry on the MiSeq FGx® Sequencing System, which is approved for National DNA Index System (NDIS) use. This study included baseline performance, sensitivity and reproducibility studies, challenging casework samples, assessments of library plexities, comparisons with capillary electrophoresis (CE), standard versus enhanced PCR1 buffers, manual versus automated library preparation, and different flow cell types. To assess reproducibility across users, 33 samples were prepared by both an experienced and a novice user. Overall, the ForenSeq™ MainstAY Kit demonstrated high reliability with pristine samples and consistent performance across users, with minimal locus dropout until 62 pg. The kit outperformed CE for low-input and degraded samples, but limitations were observed at higher plexities. Increased sample numbers on a single flow cell disproportionately reduced read counts and locus recovery in degraded casework samples, while pristine samples were less affected. Strategic adjustments, such as batching degraded samples together or lowering plexity, may help mitigate these effects. Casework samples, including blood, touched items, bone, hair, and tissue, were tested with both standard and enhanced PCR1 buffers. In a 64-sample run (32 per buffer type), increased multiplexing reduced locus recovery in standard buffer samples, while the enhanced buffer improved recovery in 16 samples. Automated library preparation on the Opentrons OT-2 produced comparable or improved recovery relative to manual preparation, with no evidence of cross-contamination. These findings provide practical guidance for forensic laboratories adopting the MainstAY Kit, particularly in optimizing workflows for challenging samples and high-throughput sequencing environments.

Journal of Forensic SciencesVol. 71(2)
Sam Houston State University (US)
Gender equality
Openalex Percentile: Top 93%
Forensic and Genetic Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Evaluation of the <scp>ForenSeq</scp> ™ <scp>MainstAY</scp> workflow for forensic <scp>DNA</scp> samples — Ryan Gutierrez, Rachel M. Houston, et al. · Journal of Forensic Sciences (2026) | TGRS Research Map | TGRS