<i>Objective.</i>In the revised version of the TRS-398 Code of Practice (CoP), Monte Carlo (MC) results were added to existing experimental data to derive the recommended beam quality correction factors (<i>k</i><sub><i>Q</i></sub>) for ionisation chambers in proton beams. While part of these results were obtained from versions v10.3 and v10.4 of the Geant4 simulation tool, this paper demonstrates that the use of a more recent version, such as v11.2, can affect the value of the<i>k</i><sub><i>Q</i></sub>factors.<i>Approach.</i>The chamber-specific proton contributions (<i>f</i><sub><i>Q</i></sub>) of the<i>k</i><sub><i>Q</i></sub>factors were derived for four ionisation chambers using two different versions of the code, namely Geant4-v.10.3 and Geant4-v11.2. A comparison of the total absorbed dose values is performed, as well as the comparison of the dose contribution for primary and secondary particles.<i>Main results.</i>Larger absorbed dose values per incident particle were derived with Geant4-v11.2 compared to Geant4-v10.3 especially for dose-to-air at high proton beam energies between 150 MeV and 250 MeV, leading to deviations in the<i>k</i><sub><i>Q</i></sub>values up to 1%. These deviations are mainly due to a change in the physics of secondary helium ions for which the significant deviations between the Geant4 versions is the most stringent within the entrance window or the shell of the ionisation chambers.<i>Significance.</i>Although significant deviations in the MC calculated<i>f</i><sub><i>Q</i></sub>values were observed between the two Geant4 versions, the dominant uncertainty of the<i>W</i><sub>air</sub>values currently allows to achieve the agreement at the<i>k</i><sub><i>Q</i></sub>level. As these values also agree with the current data presented in the TRS-398 CoP, it is not possible at the moment to discriminate between Geant4-v10.3 and Geant4-v11.2, which are therefore both suitable for<i>k</i><sub><i>Q</i></sub>calculation.

Authors

Institutions

Publication Details

Journal
Biomedical Physics & Engineering Express
Published
2026-01-08
DOI
https://doi.org/10.1088/2057-1976/ae3571
Primary Topic
Radiation Therapy and Dosimetry
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Monte Carlo derivation of beam quality correction factors in proton beams: a comparison of Geant4 versions

Kilian‐Simon Baumann, Nick Reynaert, Guillaume Gabriel Houyoux
Biomedical Physics & Engineering Express
Radiation Therapy and Dosimetry
article

Monte Carlo derivation of beam quality correction factors in proton beams: a comparison of Geant4 versions

Kilian‐Simon Baumann, Nick Reynaert, Guillaume Gabriel Houyoux
article en

Abstract

<i>Objective.</i>In the revised version of the TRS-398 Code of Practice (CoP), Monte Carlo (MC) results were added to existing experimental data to derive the recommended beam quality correction factors (<i>k</i><sub><i>Q</i></sub>) for ionisation chambers in proton beams. While part of these results were obtained from versions v10.3 and v10.4 of the Geant4 simulation tool, this paper demonstrates that the use of a more recent version, such as v11.2, can affect the value of the<i>k</i><sub><i>Q</i></sub>factors.<i>Approach.</i>The chamber-specific proton contributions (<i>f</i><sub><i>Q</i></sub>) of the<i>k</i><sub><i>Q</i></sub>factors were derived for four ionisation chambers using two different versions of the code, namely Geant4-v.10.3 and Geant4-v11.2. A comparison of the total absorbed dose values is performed, as well as the comparison of the dose contribution for primary and secondary particles.<i>Main results.</i>Larger absorbed dose values per incident particle were derived with Geant4-v11.2 compared to Geant4-v10.3 especially for dose-to-air at high proton beam energies between 150 MeV and 250 MeV, leading to deviations in the<i>k</i><sub><i>Q</i></sub>values up to 1%. These deviations are mainly due to a change in the physics of secondary helium ions for which the significant deviations between the Geant4 versions is the most stringent within the entrance window or the shell of the ionisation chambers.<i>Significance.</i>Although significant deviations in the MC calculated<i>f</i><sub><i>Q</i></sub>values were observed between the two Geant4 versions, the dominant uncertainty of the<i>W</i><sub>air</sub>values currently allows to achieve the agreement at the<i>k</i><sub><i>Q</i></sub>level. As these values also agree with the current data presented in the TRS-398 CoP, it is not possible at the moment to discriminate between Geant4-v10.3 and Geant4-v11.2, which are therefore both suitable for<i>k</i><sub><i>Q</i></sub>calculation.

Biomedical Physics & Engineering ExpressVol. 12(1)
Université Libre de Bruxelles (BE), Universitätsklinikum Gießen und Marburg (DE), Technische Hochschule Mittelhessen (DE)
Peace, Justice and strong institutions, Reduced inequalities
Openalex Percentile: Top 94%
Radiation Therapy and Dosimetry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.