Comparison of two validated oscillometric devices in a home-like setup reveals pronounced blood pressure differences and reduced precision
Home blood pressure monitoring (HBPM) is essential for long-term hypertension management, but its accuracy and reliability is questionable due to user inconsistencies and non-standard usage conditions. This study compared two validated HBPM devices - a wrist-based and a reference upper-arm monitor - in a home-like setup where participants placed the devices themselves. A total of 121 participants underwent four concurrent blood pressure (BP) measurements, with two taken in each arm, followed by a crossover of devices between arms after the initial two readings. Wrist-derived BP readings were higher than upper-arm measurements. With higher blood pressure levels, both devices exhibited greater bias, accompanied by reduced precision in systolic BP measurements. Additionally, hypertension classification showed only moderate agreement (Cohen's kappa=0.58). The wrist monitor tended to over-diagnose hypertension and exhibited greater variability than the upper-arm device. These findings highlight the need for more guidance and education as well as critical evaluation of home blood pressure measurements provided by patients.
Authors
- Brian Thompson (ORCID: https://orcid.org/0000-0003-0983-2762)
- Joachim Zahnd
- Patrick Taffé
- Pierre-Antonin Rigon
- Gregoire Wuerzner
Institutions
- Centre Hospitalier Universitaire Vaudois (CH)
- Centre universitaire de médecine générale et santé publique, Lausanne (CH)
- University of Lausanne (CH)
Publication Details
- Journal
- Hypertension Research
- Published
- 2026-01-09
- DOI
- https://doi.org/10.1038/s41440-025-02514-3
- Primary Topic
- Blood Pressure and Hypertension Studies
- Type
- article
- Field-Weighted Citation Impact
- 0.00