Pediatric dental cone-beam computed tomography using half-acquisition and low-noise reconstruction: visual evaluation of clinical images
This study evaluated whether half-acquisition (180° scan) pediatric cone-beam computed tomography (CBCT; 3D Accuitomo F17, J. Morita, Kyoto, Japan) reduces radiation exposure while maintaining sufficient diagnostic image quality for identifying ectopic eruptions and impacted teeth. Additionally, it was investigated whether a low-noise reconstruction filter (G_101) mitigates image quality degradation in 180° scans. Three board-certified oral and maxillofacial radiologists certified by the Japanese Society for Oral and Maxillofacial Radiology visually evaluated clinical images from 12 pediatric patients (aged 6-10 years). The image quality was objectively assessed using phantom-based analyses of the modulation transfer function (MTF), noise power spectrum (NPS), and comprehensive objective image quality calculated from MTF and NPS values. Although 180° images showed increased noise and slightly lower visual assessment scores compared with 360° images, they remained diagnostically acceptable. In 180° reconstructions, the median visual scores with the G_101 filter were slightly higher than those with the standard G_001 filter, with small differences (within approximately 0-3 points on a 100-point scale), although the differences were not statistically significant. Interestingly, in approximately 28% of 180 evaluations, 180° images scored higher than 360° images, likely due to reduced motion artefacts from shorter acquisition. In a previous phantom experiment, the dose area product (DAP) for 360° and 180° scans was 490 mGy cm<sup>2</sup> and 249 mGy cm<sup>2</sup>, respectively, indicating that 180° scan reduces radiation exposure while maintaining clinically acceptable image quality. These findings suggest that half-acquisition, when combined with an appropriate reconstruction filter, may offer a practical, low-dose alternative for pediatric dental imaging.
Authors
- Shu Onodera (ORCID: https://orcid.org/0000-0002-5761-6434)
- M. Ito (ORCID: https://orcid.org/0009-0004-4516-7370)
- Masahiro Iikubo (ORCID: https://orcid.org/0000-0001-9255-2141)
- K. Chida
- Masayuki Zuguchi
- Toshiki Kato (ORCID: https://orcid.org/0000-0003-1871-8064)
- Masaki Fujisawa
- Masahiro Sai
- Masaaki Nakamura (ORCID: https://orcid.org/0000-0002-5901-260X)
- Ikuho Kojima
Institutions
- Tohoku University (JP)
- Tohoku University Hospital (JP)
Publication Details
- Journal
- Physical and Engineering Sciences in Medicine
- Published
- 2026-01-08
- DOI
- https://doi.org/10.1007/s13246-025-01691-2
- Primary Topic
- Dental Radiography and Imaging
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- Japan Society for the Promotion of Science