Pediatric dental cone-beam computed tomography using half-acquisition and low-noise reconstruction: visual evaluation of clinical images

This study evaluated whether half-acquisition (180° scan) pediatric cone-beam computed tomography (CBCT; 3D Accuitomo F17, J. Morita, Kyoto, Japan) reduces radiation exposure while maintaining sufficient diagnostic image quality for identifying ectopic eruptions and impacted teeth. Additionally, it was investigated whether a low-noise reconstruction filter (G_101) mitigates image quality degradation in 180° scans. Three board-certified oral and maxillofacial radiologists certified by the Japanese Society for Oral and Maxillofacial Radiology visually evaluated clinical images from 12 pediatric patients (aged 6-10 years). The image quality was objectively assessed using phantom-based analyses of the modulation transfer function (MTF), noise power spectrum (NPS), and comprehensive objective image quality calculated from MTF and NPS values. Although 180° images showed increased noise and slightly lower visual assessment scores compared with 360° images, they remained diagnostically acceptable. In 180° reconstructions, the median visual scores with the G_101 filter were slightly higher than those with the standard G_001 filter, with small differences (within approximately 0-3 points on a 100-point scale), although the differences were not statistically significant. Interestingly, in approximately 28% of 180 evaluations, 180° images scored higher than 360° images, likely due to reduced motion artefacts from shorter acquisition. In a previous phantom experiment, the dose area product (DAP) for 360° and 180° scans was 490 mGy cm<sup>2</sup> and 249 mGy cm<sup>2</sup>, respectively, indicating that 180° scan reduces radiation exposure while maintaining clinically acceptable image quality. These findings suggest that half-acquisition, when combined with an appropriate reconstruction filter, may offer a practical, low-dose alternative for pediatric dental imaging.

Authors

Institutions

Publication Details

Journal
Physical and Engineering Sciences in Medicine
Published
2026-01-08
DOI
https://doi.org/10.1007/s13246-025-01691-2
Primary Topic
Dental Radiography and Imaging
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Pediatric dental cone-beam computed tomography using half-acquisition and low-noise reconstruction: visual evaluation of clinical images

Shu Onodera, M. Ito, Masahiro Iikubo, K. Chida et al.
Physical and Engineering Sciences in Medicine
Dental Radiography and Imaging
article

Pediatric dental cone-beam computed tomography using half-acquisition and low-noise reconstruction: visual evaluation of clinical images

Shu Onodera, M. Ito, Masahiro Iikubo, K. Chida, Masayuki Zuguchi, Toshiki Kato, Masaki Fujisawa, Masahiro Sai, Masaaki Nakamura, Ikuho Kojima
article en

Abstract

This study evaluated whether half-acquisition (180° scan) pediatric cone-beam computed tomography (CBCT; 3D Accuitomo F17, J. Morita, Kyoto, Japan) reduces radiation exposure while maintaining sufficient diagnostic image quality for identifying ectopic eruptions and impacted teeth. Additionally, it was investigated whether a low-noise reconstruction filter (G_101) mitigates image quality degradation in 180° scans. Three board-certified oral and maxillofacial radiologists certified by the Japanese Society for Oral and Maxillofacial Radiology visually evaluated clinical images from 12 pediatric patients (aged 6-10 years). The image quality was objectively assessed using phantom-based analyses of the modulation transfer function (MTF), noise power spectrum (NPS), and comprehensive objective image quality calculated from MTF and NPS values. Although 180° images showed increased noise and slightly lower visual assessment scores compared with 360° images, they remained diagnostically acceptable. In 180° reconstructions, the median visual scores with the G_101 filter were slightly higher than those with the standard G_001 filter, with small differences (within approximately 0-3 points on a 100-point scale), although the differences were not statistically significant. Interestingly, in approximately 28% of 180 evaluations, 180° images scored higher than 360° images, likely due to reduced motion artefacts from shorter acquisition. In a previous phantom experiment, the dose area product (DAP) for 360° and 180° scans was 490 mGy cm<sup>2</sup> and 249 mGy cm<sup>2</sup>, respectively, indicating that 180° scan reduces radiation exposure while maintaining clinically acceptable image quality. These findings suggest that half-acquisition, when combined with an appropriate reconstruction filter, may offer a practical, low-dose alternative for pediatric dental imaging.

Physical and Engineering Sciences in Medicine
Tohoku University (JP), Tohoku University Hospital (JP)
Japan Society for the Promotion of Science
Openalex Percentile: Top 93%
Dental Radiography and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.