Validation of SRS MapCHECK for CyberKnife patient-specific quality assurance: challenges with small cone sizes
This study evaluates the suitability of SRS MapCHECK for PSQA of the CyberKnife system and highlights its limitations. It demonstrates that while SRS MapCHECK is effective for larger fixed cone sizes, it does not ensure dosimetric accuracy for plans involving very small fixed cones.
Authors
- Hojae Kim
- Hojin Kim
- Jin Sung Kim
- Dohyeon Yoo
- Taeho Kim
Institutions
- Yonsei University (KR)
Publication Details
- Journal
- Frontiers in Oncology
- Published
- 2026-01-08
- DOI
- https://doi.org/10.3389/fonc.2025.1667108
- Primary Topic
- Advanced Radiotherapy Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- Korea Medical Device Development Fund