Pole-To-Pole 3D Radial Trajectory Designs Improve Image Quality and Quantitative Parametric Mapping in the Brain and Heart.

Purpose To design 3D radial spiral phyllotaxis trajectories aimed at removing phase inconsistencies, improving image quality, and enhancing parametric mapping accuracy by acquiring nearly opposing spokes starting from both hemispheres in 3D radial k-space.Methods Two 3D radial trajectories, pole-to-pole and continuous spiral phyllotaxis, were developed and implemented on a 3T MRI scanner in a phase-cycled balanced steady-state free precession (bSSFP) and a spoiled gradient-echo (GRE) sequence. Image quality and k-space center phase variations were evaluated in a spherical phantom using the original and new radial phyllotaxis designs. T1/T2 was quantified and compared using phase-cycled bSSFP data acquired with the new radial trajectory designs, as well as the original phyllotaxis trajectory and a Cartesian trajectory as references, in both an MRI system phantom and the brains of three healthy volunteers. ECG-triggered whole-heart GRE data were acquired using the original and pole-to-pole phyllotaxis trajectories in three healthy volunteers and compared for image quality improvement.Results All 3D radial trajectory designs showed variations in the k-space center phase depending on the orientation of the readout spokes. Image quality improved when using the pole-to-pole and continuous phyllotaxis over the original trajectory. Scans using the original trajectory had higher T1/T2 estimation errors in comparison to the new trajectories and the Cartesian trajectory. The pole-to-pole and continuous trajectories improved T1/T2 maps of the brain and image quality for all cardiac images.Conclusion Acquiring nearly opposing spokes in 3D radial trajectory designs compensates phase inconsistencies without requiring additional corrections, which improves quantitative imaging and anatomical visualizations.

Authors

Institutions

Publication Details

Journal
Open Access CRIS of the University of Bern
Published
2026-01-06
DOI
https://doi.org/10.48620/93623
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Pole-To-Pole 3D Radial Trajectory Designs Improve Image Quality and Quantitative Parametric Mapping in the Brain and Heart.

Grzegorz Bauman, Yasaman Safarkhanlo, Joseph G. Woods, Oliver Bieri et al.
Open Access CRIS of the University of Bern
Advanced MRI Techniques and Applications
article

Pole-To-Pole 3D Radial Trajectory Designs Improve Image Quality and Quantitative Parametric Mapping in the Brain and Heart.

Grzegorz Bauman, Yasaman Safarkhanlo, Joseph G. Woods, Oliver Bieri, Davide Piccini, Adèle Mackowiak, Christopher Roy, Nils Plähn, J. Bastiaansen, Matteo Tagliabue, Eva S. Peper, Berk C. Açikgöz, Li Feng
article en

Abstract

Purpose To design 3D radial spiral phyllotaxis trajectories aimed at removing phase inconsistencies, improving image quality, and enhancing parametric mapping accuracy by acquiring nearly opposing spokes starting from both hemispheres in 3D radial k-space.Methods Two 3D radial trajectories, pole-to-pole and continuous spiral phyllotaxis, were developed and implemented on a 3T MRI scanner in a phase-cycled balanced steady-state free precession (bSSFP) and a spoiled gradient-echo (GRE) sequence. Image quality and k-space center phase variations were evaluated in a spherical phantom using the original and new radial phyllotaxis designs. T1/T2 was quantified and compared using phase-cycled bSSFP data acquired with the new radial trajectory designs, as well as the original phyllotaxis trajectory and a Cartesian trajectory as references, in both an MRI system phantom and the brains of three healthy volunteers. ECG-triggered whole-heart GRE data were acquired using the original and pole-to-pole phyllotaxis trajectories in three healthy volunteers and compared for image quality improvement.Results All 3D radial trajectory designs showed variations in the k-space center phase depending on the orientation of the readout spokes. Image quality improved when using the pole-to-pole and continuous phyllotaxis over the original trajectory. Scans using the original trajectory had higher T1/T2 estimation errors in comparison to the new trajectories and the Cartesian trajectory. The pole-to-pole and continuous trajectories improved T1/T2 maps of the brain and image quality for all cardiac images.Conclusion Acquiring nearly opposing spokes in 3D radial trajectory designs compensates phase inconsistencies without requiring additional corrections, which improves quantitative imaging and anatomical visualizations.

Open Access CRIS of the University of Bern
Society of Interventional Radiology (US)
Openalex Percentile: Top 95%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.