Reliability and Device Comparison of a Newly Developed Surface Electromyography Device: A Preliminary Study on Healthy Adults
Purpose: This study aimed to evaluate the reliability of a newly developed wireless surface electromyography (sEMG) device by comparing it to an existing commercial sEMG system in healthy adults.Methods: Twenty healthy adults participated in the study.The sEMG signals were recorded using a newly developed wireless sEMG device (K-sEMG40) and a commercial system (BTS-FREEEMG1000) during maximal voluntary isometric contraction (MVIC) and submaximal contractions (40% and 80% MVIC).Reliability and agreement were assessed using intraclass correlation coefficients (ICC), Spearman's correlation, Bland-Altman analysis, and effect size calculation. Results:The developed device showed high reliability (ICC > 0.85) for MVIC and 40% MVIC, with moderate reliability at 80% MVIC.Strong correlations (r > 0.86, p < 0.01) were found between the two devices.The Bland-Altman analysis demonstrated good agreement, and the effect size analysis indicated small differences between measurements. Conclusion:The newly developed wireless sEMG device demonstrated high reliability and agreement with the existing
Authors
- Jong-Ho Kang (ORCID: https://orcid.org/0000-0002-6220-7272)
- Taesung Park (ORCID: https://orcid.org/0000-0003-1625-6062)
- Myung-Jun Shin
- Bu-Young Choi
Institutions
- Catholic University of Pusan (KR)
- Pusan National University Hospital (KR)
- Pusan National University Yangsan Hospital (KR)
- Pusan National University (KR)
Publication Details
- Journal
- PNF and Movement
- Published
- 2025-04-30
- DOI
- https://doi.org/10.21598/jkpnfa.2025.23.1.79
- Primary Topic
- Muscle activation and electromyography studies
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- Pusan National University