Dynamic testing of cortisol in consecutive patients with suspected adrenal disorders: a comparison of Elecsys Cortisol II and LC-MS/MS

The increased detection of adrenal incidentalomas and drug-induced adrenal insufficiency increases the demand for cortisol testing. While liquid chromatography-tandem mass spectrometry (LC-MS/MS) and dynamic testing remain the gold standard, more expedient tools are needed. We compared the Elecsys Cortisol II immunoassay (ElecsysCort II) with LC-MS/MS during dynamic testing and evaluated the diagnostic performance of baseline cortisol cutoffs (0 min cortisol during a short Synacthen test). The study included 547 overnight dexamethasone suppression tests (44% abnormal), mainly performed in adrenal incidentaloma patients (69%), and 519 Synacthen tests (32% abnormal). ElecsysCort II slightly underestimated cortisol compared with LC-MS/MS with a Passing-Bablok regression of -3.20 (95% CI: -3.66 to -2.83) + 0.96X (95% CI: 0.96-0.97) and a mean relative difference of -8.22% (95% CI: -8.99% to -7.45%) by Bland-Altman, which was within the limits of acceptable bias based on biological variation. Agreement was not affected by glucocorticoid or estradiol intake. Compared with LC-MS/MS, ElecsysCort II demonstrated a specificity of 100%, sensitivity of 84%, positive predictive value of 99%, and negative predictive value of 86% for an abnormal overnight dexamethasone suppression test (480 min cortisol ≥50 nmol/L). For an abnormal Synacthen test (30 min cortisol <420 nmol/L), the measures were 95, 99, 89, and 100%, respectively. A baseline cortisol cutoff >300 nmol/L to rule out and <150 nmol/L to rule in adrenal insufficiency demonstrated a high specificity (>92%) and a positive predictive value (>87%). Our study supports the use of ElecsysCort II in general and baseline cortisol as a screening tool for adrenal insufficiency. The high proportion of abnormal overnight dexamethasone suppression tests among incidentaloma patients warrants further research.

Authors

Institutions

Publication Details

Journal
Endocrine Connections
Published
2026-01-01
DOI
https://doi.org/10.1530/ec-25-0600
Primary Topic
Adrenal Hormones and Disorders
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Dynamic testing of cortisol in consecutive patients with suspected adrenal disorders: a comparison of Elecsys Cortisol II and LC-MS/MS

Simon Bøggild Hansen, Søren Ladefoged, Jens Otto Lunde Jørgensen, Kristina Laugesen et al.
Endocrine Connections
Adrenal Hormones and Disorders
article

Dynamic testing of cortisol in consecutive patients with suspected adrenal disorders: a comparison of Elecsys Cortisol II and LC-MS/MS

Simon Bøggild Hansen, Søren Ladefoged, Jens Otto Lunde Jørgensen, Kristina Laugesen, Lise Sofie Bislev
article en

Abstract

The increased detection of adrenal incidentalomas and drug-induced adrenal insufficiency increases the demand for cortisol testing. While liquid chromatography-tandem mass spectrometry (LC-MS/MS) and dynamic testing remain the gold standard, more expedient tools are needed. We compared the Elecsys Cortisol II immunoassay (ElecsysCort II) with LC-MS/MS during dynamic testing and evaluated the diagnostic performance of baseline cortisol cutoffs (0 min cortisol during a short Synacthen test). The study included 547 overnight dexamethasone suppression tests (44% abnormal), mainly performed in adrenal incidentaloma patients (69%), and 519 Synacthen tests (32% abnormal). ElecsysCort II slightly underestimated cortisol compared with LC-MS/MS with a Passing-Bablok regression of -3.20 (95% CI: -3.66 to -2.83) + 0.96X (95% CI: 0.96-0.97) and a mean relative difference of -8.22% (95% CI: -8.99% to -7.45%) by Bland-Altman, which was within the limits of acceptable bias based on biological variation. Agreement was not affected by glucocorticoid or estradiol intake. Compared with LC-MS/MS, ElecsysCort II demonstrated a specificity of 100%, sensitivity of 84%, positive predictive value of 99%, and negative predictive value of 86% for an abnormal overnight dexamethasone suppression test (480 min cortisol ≥50 nmol/L). For an abnormal Synacthen test (30 min cortisol <420 nmol/L), the measures were 95, 99, 89, and 100%, respectively. A baseline cortisol cutoff >300 nmol/L to rule out and <150 nmol/L to rule in adrenal insufficiency demonstrated a high specificity (>92%) and a positive predictive value (>87%). Our study supports the use of ElecsysCort II in general and baseline cortisol as a screening tool for adrenal insufficiency. The high proportion of abnormal overnight dexamethasone suppression tests among incidentaloma patients warrants further research.

Endocrine ConnectionsVol. 15(1)
Aarhus University Hospital (DK), Regional Hospital Horsens (DK), Regional Hospital Randers (DK)
Good health and well-being
Openalex Percentile: Top 95%
Adrenal Hormones and Disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.