Improved Quantification of ICG Perfusion Through Motion Compensation in Fluorescence-Guided Surgery

<b>Background/Objectives</b>: Motion artifacts significantly distort fluorescence measurements during surgical perfusion assessment, potentially leading to incorrect clinical decisions. This study evaluates the efficacy of automated motion compensation (MC) in quantitative indocyanine green (q-ICG) imaging to improve the accuracy of perfusion assessment. <b>Methods</b>: Frames from ICG perfusion assessment during 17 pancreaticoduodenectomies were analyzed. Regions of interest (ROIs) were systematically placed on each frame series, and automated MC was applied to track tissue movement. Performance was evaluated by comparing MC with surgeon-adjusted placement using multiple image quality metrics and analyzing perfusion metrics on time-intensity curves. Principal Component Analysis (PCA) was applied to explore whether image patterns could distinguish between successful and unsuccessful motion compensation. <b>Results</b>: Automated motion compensation successfully corrected motion artifacts in 67.5% of frame sequences, achieving comparable performance to surgeon-guided adjustments. PCA demonstrated clear separation between sufficient and insufficient corrections (AUC = 0.80). At the population level, MC did not significantly change perfusion slope (<i>t</i>(59) = 1.60, <i>p</i> = 0.11) or time-to-peak (Tmax; <i>t</i>(58) = 0.81, <i>p</i> = 0.42). Bland-Altman analysis showed a mean bias of -0.54 (SD = 3.32) for slope and 24.95 (SD = 238.40) for Tmax. At the individual level, 86.7% of slope and 79.7% of Tmax values differed by ≥10% after MC, with mean absolute percentage changes of 108.5% (median 37.8%) and 431.5% (median 65.9%), respectively. <b>Conclusions</b>: MC effectively reduces motion artifacts in fluorescence-guided perfusion assessment. By improving the precision of ICG-derived parameters, this technology enhances measurement reliability and represents an enabler for accurate intraoperative perfusion quantification.

Authors

Institutions

Publication Details

Journal
Diagnostics
Published
2026-01-06
DOI
https://doi.org/10.3390/diagnostics16020176
Primary Topic
Optical Imaging and Spectroscopy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Improved Quantification of ICG Perfusion Through Motion Compensation in Fluorescence-Guided Surgery

Nikolaj Nerup, Morten Bo Søndergaard Svendsen, Thomas B. Piper, Sermed Nicolae et al.
Diagnostics
Optical Imaging and Spectroscopy Techniques
article

Improved Quantification of ICG Perfusion Through Motion Compensation in Fluorescence-Guided Surgery

Nikolaj Nerup, Morten Bo Søndergaard Svendsen, Thomas B. Piper, Sermed Nicolae, Michael Patrick Achiam
article en

Abstract

<b>Background/Objectives</b>: Motion artifacts significantly distort fluorescence measurements during surgical perfusion assessment, potentially leading to incorrect clinical decisions. This study evaluates the efficacy of automated motion compensation (MC) in quantitative indocyanine green (q-ICG) imaging to improve the accuracy of perfusion assessment. <b>Methods</b>: Frames from ICG perfusion assessment during 17 pancreaticoduodenectomies were analyzed. Regions of interest (ROIs) were systematically placed on each frame series, and automated MC was applied to track tissue movement. Performance was evaluated by comparing MC with surgeon-adjusted placement using multiple image quality metrics and analyzing perfusion metrics on time-intensity curves. Principal Component Analysis (PCA) was applied to explore whether image patterns could distinguish between successful and unsuccessful motion compensation. <b>Results</b>: Automated motion compensation successfully corrected motion artifacts in 67.5% of frame sequences, achieving comparable performance to surgeon-guided adjustments. PCA demonstrated clear separation between sufficient and insufficient corrections (AUC = 0.80). At the population level, MC did not significantly change perfusion slope (<i>t</i>(59) = 1.60, <i>p</i> = 0.11) or time-to-peak (Tmax; <i>t</i>(58) = 0.81, <i>p</i> = 0.42). Bland-Altman analysis showed a mean bias of -0.54 (SD = 3.32) for slope and 24.95 (SD = 238.40) for Tmax. At the individual level, 86.7% of slope and 79.7% of Tmax values differed by ≥10% after MC, with mean absolute percentage changes of 108.5% (median 37.8%) and 431.5% (median 65.9%), respectively. <b>Conclusions</b>: MC effectively reduces motion artifacts in fluorescence-guided perfusion assessment. By improving the precision of ICG-derived parameters, this technology enhances measurement reliability and represents an enabler for accurate intraoperative perfusion quantification.

DiagnosticsVol. 16(2)
University of Copenhagen (DK), Copenhagen University Hospital (DK), Technical University of Denmark (DK)
Openalex Percentile: Top 96%
Optical Imaging and Spectroscopy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.