Impact of icterus on PT and APTT measured with photometric versus mechanical clot detection

Abstract Objectives Icteric plasma samples, common in liver diseases, may interfere with photometric clot detection. This study was aimed to compare prothrombin time (PT) and activated partial thrompoblastin time (APTT) from non-icteric and icteric plasma samples using Werfen ACL TOP 750 and Stago sthemO 301. Methods We analyzed 100 routine plasma samples (50 icteric, 50 non-icteric) for PT and APTT using ACL TOP 750 (photometric clot detection) and sthemO 301 (mechanical clot detection). PT and APTT were measured in duplicate, and the icteric index (I-index) was recorded. Inter-analyzer agreement was assessed with correlation analysis, linear fit, bootstrap analysis and hypothesis testing for equality of regression slopes and intercepts. Results Non-icteric samples had a mean I-index of 1.0 ± 0.0 vs. 2.4 ± 0.7 in icteric samples (p<0.001). PT values showed strong inter-analyzer agreement (r=0.940 and 0.955 in non-icteric and icteric samples), while APTT exhibited moderate agreement (r=0.740 and 0.800 in non-icteric and icteric samples). Bootstrap simulations showed no significant difference in correlation coefficients between icteric and non-icteric samples. Comparison of linear fitting revealed significant differences in slopes and intercepts between non-icteric and icteric sample for both PT (slopes: p<0.001; intercepts: p=0.001) and APTT (slopes: p=0.002; intercepts: p=0.003). Conclusions Icteric plasma was found to modify the relationship between PT and APTT values measured by the two coagulation analyzers, indicating that values obtained with photometric and mechanical clot detection methods in icteric samples may not be directly interchangeable.

Authors

Institutions

Publication Details

Journal
Journal of Laboratory Medicine
Published
2026-01-06
DOI
https://doi.org/10.1515/labmed-2025-0215
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Impact of icterus on PT and APTT measured with photometric versus mechanical clot detection

Nicola Baratto, Alessandro Lorenzetto, Gian Luca Salvagno, Giuseppe Lippi
Journal of Laboratory Medicine
Clinical Laboratory Practices and Quality Control
article

Impact of icterus on PT and APTT measured with photometric versus mechanical clot detection

Nicola Baratto, Alessandro Lorenzetto, Gian Luca Salvagno, Giuseppe Lippi
article en

Abstract

Abstract Objectives Icteric plasma samples, common in liver diseases, may interfere with photometric clot detection. This study was aimed to compare prothrombin time (PT) and activated partial thrompoblastin time (APTT) from non-icteric and icteric plasma samples using Werfen ACL TOP 750 and Stago sthemO 301. Methods We analyzed 100 routine plasma samples (50 icteric, 50 non-icteric) for PT and APTT using ACL TOP 750 (photometric clot detection) and sthemO 301 (mechanical clot detection). PT and APTT were measured in duplicate, and the icteric index (I-index) was recorded. Inter-analyzer agreement was assessed with correlation analysis, linear fit, bootstrap analysis and hypothesis testing for equality of regression slopes and intercepts. Results Non-icteric samples had a mean I-index of 1.0 ± 0.0 vs. 2.4 ± 0.7 in icteric samples (p<0.001). PT values showed strong inter-analyzer agreement (r=0.940 and 0.955 in non-icteric and icteric samples), while APTT exhibited moderate agreement (r=0.740 and 0.800 in non-icteric and icteric samples). Bootstrap simulations showed no significant difference in correlation coefficients between icteric and non-icteric samples. Comparison of linear fitting revealed significant differences in slopes and intercepts between non-icteric and icteric sample for both PT (slopes: p<0.001; intercepts: p=0.001) and APTT (slopes: p=0.002; intercepts: p=0.003). Conclusions Icteric plasma was found to modify the relationship between PT and APTT values measured by the two coagulation analyzers, indicating that values obtained with photometric and mechanical clot detection methods in icteric samples may not be directly interchangeable.

Journal of Laboratory MedicineVol. 50(1)
University of Verona (IT)
Openalex Percentile: Top 96%
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.