Blood-mimicking dye phantoms for evaluating photoacoustic oximetry accuracy

Many proposed clinical applications of photoacoustic imaging (PAI) rely on relative or absolute measurements of blood oxygen saturation (sO2), and evaluation of oximetry measurement accuracy is crucial for assessing device performance. Available bench test methods use phantoms connected to blood flow circuits with tunable oxygenation, but these methods are complex, costly, and pose biohazard safety risks. To address these issues, we have developed stable and tunable blood-mimicking solutions using binary mixtures of commercially available near-infrared organic dyes (NIR746A and IRA980) to enable non-biological phantom-based PAI oximetry test methods. We used spectrophotometry and a custom PA spectroscopy system to characterize dye extinction and PA response at 750 nm and 850 nm, then formulated various dye recipes mimicking sO2 levels from 40 % to 100 %. We then used a custom PAI system to image breast-mimicking polyacrylamide hydrogel phantoms with embedded tubes injected with static volumes of either dye solutions or bovine blood deoxygenated using sodium dithionite. Phantom testing with dyes produced similar performance metrics to blood, with root-mean-squared difference (RMSD) values between photoacoustic sO2 and reference sO2 of 6-17 % for blood and 4-18 % for dyes, sensitivity (slope of the regression line) ranged from 0.4 to 0.7 for blood and 0.4-0.9 for dyes, and depth-averaged bias ranged from 4 % to 17 % for blood and 3-10 % for dyes. These blood-mimicking dyes may offer a simpler, cheaper, safer, and more stable approach to evaluate PAI oximetry accuracy compared to traditional blood flow phantoms. This tool could facilitate establishment of less burdensome and more reproducible phantom-based PAI test methods, ultimately expediting clinical adoption of PAI technology.

Authors

Institutions

Publication Details

Journal
Photoacoustics
Published
2026-01-08
DOI
https://doi.org/10.1016/j.pacs.2026.100795
Primary Topic
Photoacoustic and Ultrasonic Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Blood-mimicking dye phantoms for evaluating photoacoustic oximetry accuracy

Jesse V. Jokerst, Keith A. Wear, William C. Vogt, Yong Zhou et al.
Photoacoustics
Photoacoustic and Ultrasonic Imaging
article

Blood-mimicking dye phantoms for evaluating photoacoustic oximetry accuracy

Jesse V. Jokerst, Keith A. Wear, William C. Vogt, Yong Zhou, Zixin Wang, T. Joshua Pfefer
article en

Abstract

Many proposed clinical applications of photoacoustic imaging (PAI) rely on relative or absolute measurements of blood oxygen saturation (sO2), and evaluation of oximetry measurement accuracy is crucial for assessing device performance. Available bench test methods use phantoms connected to blood flow circuits with tunable oxygenation, but these methods are complex, costly, and pose biohazard safety risks. To address these issues, we have developed stable and tunable blood-mimicking solutions using binary mixtures of commercially available near-infrared organic dyes (NIR746A and IRA980) to enable non-biological phantom-based PAI oximetry test methods. We used spectrophotometry and a custom PA spectroscopy system to characterize dye extinction and PA response at 750 nm and 850 nm, then formulated various dye recipes mimicking sO2 levels from 40 % to 100 %. We then used a custom PAI system to image breast-mimicking polyacrylamide hydrogel phantoms with embedded tubes injected with static volumes of either dye solutions or bovine blood deoxygenated using sodium dithionite. Phantom testing with dyes produced similar performance metrics to blood, with root-mean-squared difference (RMSD) values between photoacoustic sO2 and reference sO2 of 6-17 % for blood and 4-18 % for dyes, sensitivity (slope of the regression line) ranged from 0.4 to 0.7 for blood and 0.4-0.9 for dyes, and depth-averaged bias ranged from 4 % to 17 % for blood and 3-10 % for dyes. These blood-mimicking dyes may offer a simpler, cheaper, safer, and more stable approach to evaluate PAI oximetry accuracy compared to traditional blood flow phantoms. This tool could facilitate establishment of less burdensome and more reproducible phantom-based PAI test methods, ultimately expediting clinical adoption of PAI technology.

PhotoacousticsVol. 47
Center for Devices and Radiological Health (US), University of California San Diego (US)
Openalex Percentile: Top 95%
Photoacoustic and Ultrasonic Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.