Experimental comparison between forward and reverse Vibrant Soundbridge cochlear stimulation
Forward couplings, including direct stapes coupling, are mechanically superior to reverse stimulation. The stapes head coupler performs 7-25 dB better than the round window coupler and similar or even superior to incus couplers.