Comparative Performance of a Routine Immunoturbidimetric Glycated Hemoglobin Assay and High-Pressure Liquid Chromatography Method

Introduction Hemoglobin A1c (HbA1c) testing is used for the diagnosis of diabetes and glycemic monitoring. This study compares an immunoturbidimetric assay with high-performance liquid chromatography (HPLC) to assess analytical agreement and determine the clinical suitability of the immunoturbidimetric method for diagnosis versus long-term monitoring in a local setting, in alignment with international standards. Methods A comparative cross-sectional study was conducted in a routine laboratory in Madagascar. Samples were tested on the same day using the Bio-Rad D-10 (Bio-Rad Laboratories, Hercules, CA) and Mindray BS-300 (Mindray, Shenzhen, China) automated analyzers. Analytical accuracy was assessed using control materials. The two methods were compared by evaluating the correlation using Pearson's coefficient (r) and linear regression, and the concordance using the Bland-Altman plot and Passing-Bablok regression. The HbA1c results were divided into a total group and subgroups (HbA1c < 6.5%, 6.5-8%, and > 8%). Results Intra-assay, inter-assay, and overall precision are satisfactory (coefficients of variation (CV) <3%) for routine use of both techniques. A good overall correlation between the immunoturbidimetric technique and HPLC was found with r = 0.88. The Bland-Altman plot shows a proportional increase in mean bias with HbA1c values and the regression equation [Formula: see text]. Conclusion This study showed good correlation between the two techniques, but a systematic underestimation of HbA1c levels by the immunoturbidimetry assay was observed, especially for high values.

Authors

Institutions

Publication Details

Journal
Cureus
Published
2026-01-07
DOI
https://doi.org/10.7759/cureus.101019
Primary Topic
Diabetes, Cardiovascular Risks, and Lipoproteins
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparative Performance of a Routine Immunoturbidimetric Glycated Hemoglobin Assay and High-Pressure Liquid Chromatography Method

Soja M Rakotomalala, Miora K Ranaivosoa, Faralahy H Rakotonjafiniarivo, Tokinomenjanahary Antsonantenaina
Cureus
Diabetes, Cardiovascular Risks, and Lipoproteins
article

Comparative Performance of a Routine Immunoturbidimetric Glycated Hemoglobin Assay and High-Pressure Liquid Chromatography Method

Soja M Rakotomalala, Miora K Ranaivosoa, Faralahy H Rakotonjafiniarivo, Tokinomenjanahary Antsonantenaina
article en

Abstract

Introduction Hemoglobin A1c (HbA1c) testing is used for the diagnosis of diabetes and glycemic monitoring. This study compares an immunoturbidimetric assay with high-performance liquid chromatography (HPLC) to assess analytical agreement and determine the clinical suitability of the immunoturbidimetric method for diagnosis versus long-term monitoring in a local setting, in alignment with international standards. Methods A comparative cross-sectional study was conducted in a routine laboratory in Madagascar. Samples were tested on the same day using the Bio-Rad D-10 (Bio-Rad Laboratories, Hercules, CA) and Mindray BS-300 (Mindray, Shenzhen, China) automated analyzers. Analytical accuracy was assessed using control materials. The two methods were compared by evaluating the correlation using Pearson's coefficient (r) and linear regression, and the concordance using the Bland-Altman plot and Passing-Bablok regression. The HbA1c results were divided into a total group and subgroups (HbA1c < 6.5%, 6.5-8%, and > 8%). Results Intra-assay, inter-assay, and overall precision are satisfactory (coefficients of variation (CV) <3%) for routine use of both techniques. A good overall correlation between the immunoturbidimetric technique and HPLC was found with r = 0.88. The Bland-Altman plot shows a proportional increase in mean bias with HbA1c values and the regression equation [Formula: see text]. Conclusion This study showed good correlation between the two techniques, but a systematic underestimation of HbA1c levels by the immunoturbidimetry assay was observed, especially for high values.

CureusVol. 18(1)
University of Antananarivo (MG)
Partnerships for the goals
Openalex Percentile: Top 96%
Diabetes, Cardiovascular Risks, and Lipoproteins
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.