Immediate aortic obstruction in young infants undergoing patent arterial duct closure with the duct occluder

We report three young infants (118, 45, and 80 days; 3.5, 4.1, and 3.8 kg) with patent ductus arteriosus (PDA) who had a uniform pattern of immediate device displacement into the aorta after an apparently secure deployment of a duct occluder. All had aortic obstruction requiring urgent surgical relief. During the study period, 18 infants (<6 months) underwent PDA closure with the same device size (6-4 Lifetech duct occluder). The aortic diameter and PDA length were similar in cases versus the 15 controls with uneventful procedures. The percentage shortening screw to pin distance ratio, which indicates the degree of recoil of the device upon release, was significantly higher in the cases versus controls (14.00 [12-15] vs. 4.00 [2.30-7.50], <i>P</i> = 0.011). A combination of tissue pliability and exaggerated recoil can occasionally result in unforeseen device migration and resultant aortic obstruction in small young infants, meriting cautious assessment before deployment and immediately after release.

Authors

Institutions

Publication Details

Journal
Annals of Pediatric Cardiology
Published
2025-09-01
DOI
https://doi.org/10.4103/apc.apc_228_25
Citations
1
Primary Topic
Cardiovascular Conditions and Treatments
Type
article
Field-Weighted Citation Impact
1.36
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Immediate aortic obstruction in young infants undergoing patent arterial duct closure with the duct occluder

Taniya Rachel Issacs, Navaneetha Sasikumar, Raman Krishna Kumar
1 citations
Annals of Pediatric Cardiology
Cardiovascular Conditions and Treatments
1.36
article

Immediate aortic obstruction in young infants undergoing patent arterial duct closure with the duct occluder

Taniya Rachel Issacs, Navaneetha Sasikumar, Raman Krishna Kumar
article en
1 citations

Abstract

We report three young infants (118, 45, and 80 days; 3.5, 4.1, and 3.8 kg) with patent ductus arteriosus (PDA) who had a uniform pattern of immediate device displacement into the aorta after an apparently secure deployment of a duct occluder. All had aortic obstruction requiring urgent surgical relief. During the study period, 18 infants (<6 months) underwent PDA closure with the same device size (6-4 Lifetech duct occluder). The aortic diameter and PDA length were similar in cases versus the 15 controls with uneventful procedures. The percentage shortening screw to pin distance ratio, which indicates the degree of recoil of the device upon release, was significantly higher in the cases versus controls (14.00 [12-15] vs. 4.00 [2.30-7.50], <i>P</i> = 0.011). A combination of tissue pliability and exaggerated recoil can occasionally result in unforeseen device migration and resultant aortic obstruction in small young infants, meriting cautious assessment before deployment and immediately after release.

Annals of Pediatric CardiologyVol. 18(5)
Amrita Institute of Medical Sciences and Research Centre (IN)
Openalex Percentile: Top 13%
Cardiovascular Conditions and Treatments
1.36
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Immediate aortic obstruction in young infants undergoing patent arterial duct closure with the duct occluder — Taniya Rachel Issacs, Navaneetha Sasikumar, et al. · Annals of Pediatric Cardiology (2025) | TGRS Research Map | TGRS