Agreement and repeatability of intraocular pressure measurements between SK-5500B and Icare IC200 tonometers in sitting and supine positions: a comparative study
Despite wide limits of agreement in Bland-Altman analysis, the SK-5500B showed good agreement and highly comparable repeatability with the Icare IC200 in both sitting and supine positions, supporting its potential as a practical clinical alternative.