Computed tomography analysis of acetabular component anteversion after total hip arthroplasty: A retrospective study comparing conventional axial and functional coronal planes

This study aims to compare the accuracy and reliability of computed tomography (CT) anteversion measurements between the conventional axial plane (CAP) and the functional coronal plane (FCP). We retrospectively examined 68 patients' postoperative CT scans following primary unilateral total hip arthroplasty (THA). PolyWare software, which uses edge detection to determine radiographic anteversion of the acetabular component, served as the reference standard. We measured anatomical anteversion in the CAP and radiographic anteversion in the reformatted FCP using reconstruction software. Reliability was evaluated with the intraclass correlation coefficient. The mean anteversion of the acetabular component in the CAP and FCP CT images were 29.18 ± 6.06° and 21.17 ± 5.38°, respectively. The mean CT anteversion in the FCP was considered accurate owing to its similarity with the reference standard (20.70 ± 4.39°, P = .281). However, the mean CT anteversion in the CAP was significantly different and thus considered less accurate; the CT anteversion in the CAP was larger than that in the FCP, showing a significant linear correlation (R = 0.673, P < .001), with a difference of 8.01°. The intraobserver and interobserver reliabilities were excellent for all measurement methods. Using radiographic anteversion defined by digital image analysis software as the reference standard, the anteversion measurement in the reformatted FCP proved more accurate than in the CAP. The reformatted FCP CT image facilitates precise post-THA anteversion measurement, with the method difference being around 8.01°. This study aids in measuring anteversion via CT.

Authors

Institutions

Publication Details

Journal
Medicine
Published
2026-01-02
DOI
https://doi.org/10.1097/md.0000000000046790
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Computed tomography analysis of acetabular component anteversion after total hip arthroplasty: A retrospective study comparing conventional axial and functional coronal planes

Kuen Tak Suh, Jung Sub Lee, Suk-Woong Kang, Won Chul Shin et al.
Medicine
Orthopaedic implants and arthroplasty
article

Computed tomography analysis of acetabular component anteversion after total hip arthroplasty: A retrospective study comparing conventional axial and functional coronal planes

Kuen Tak Suh, Jung Sub Lee, Suk-Woong Kang, Won Chul Shin, Min Uk Do, Sang-Min Lee
article en

Abstract

This study aims to compare the accuracy and reliability of computed tomography (CT) anteversion measurements between the conventional axial plane (CAP) and the functional coronal plane (FCP). We retrospectively examined 68 patients' postoperative CT scans following primary unilateral total hip arthroplasty (THA). PolyWare software, which uses edge detection to determine radiographic anteversion of the acetabular component, served as the reference standard. We measured anatomical anteversion in the CAP and radiographic anteversion in the reformatted FCP using reconstruction software. Reliability was evaluated with the intraclass correlation coefficient. The mean anteversion of the acetabular component in the CAP and FCP CT images were 29.18 ± 6.06° and 21.17 ± 5.38°, respectively. The mean CT anteversion in the FCP was considered accurate owing to its similarity with the reference standard (20.70 ± 4.39°, P = .281). However, the mean CT anteversion in the CAP was significantly different and thus considered less accurate; the CT anteversion in the CAP was larger than that in the FCP, showing a significant linear correlation (R = 0.673, P < .001), with a difference of 8.01°. The intraobserver and interobserver reliabilities were excellent for all measurement methods. Using radiographic anteversion defined by digital image analysis software as the reference standard, the anteversion measurement in the reformatted FCP proved more accurate than in the CAP. The reformatted FCP CT image facilitates precise post-THA anteversion measurement, with the method difference being around 8.01°. This study aids in measuring anteversion via CT.

MedicineVol. 105(1)
Scripps Health (US), Pusan National University Hospital (KR), Scripps Clinic (US), Pusan National University Yangsan Hospital (KR), Pusan National University (KR)
Openalex Percentile: Top 96%
Orthopaedic implants and arthroplasty
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.