Quantitative determination of al (III) traces in soft drink, pharmaceutical products and biological fluids of kidney failure and alzheimer disease patients using carbon sensor

Carbon paste sensor based on p-chlorophenyl maleanilic acid (pCl-MA) as an ionophore was synthesized and investigated as an Al(III) ion selective sensor. The developed sensor shows Nernstian slope of 20.32 ± 1.18 mV decade− 1 over the working concentration range of 1.0 × 10− 6 − 1.0 × 10− 1 mol L− 1 of Al (III) ion with a detection limit of 3.3 × 10− 7 mol L− 1. The sensor exhibited a stable potential response in the pH range from 4.0 to 6.0 and a fast response time of 8 s. The prepared sensor indicated a very good selectivity to Al (III) over a variety of cations. The sensor could be easily employed for determination of Al (III) ions in pharmaceutical products, soft drink and biological fluids of kidney failure and Alzheimer disease patients. The reaction mechanism between pCL-MA and Al (III) was confirmed by FT-IR spectroscopic analysis, SEM and EDX.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2026-01-06
DOI
https://doi.org/10.1038/s41598-025-32308-z
Primary Topic
Analytical Chemistry and Sensors
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Quantitative determination of al (III) traces in soft drink, pharmaceutical products and biological fluids of kidney failure and alzheimer disease patients using carbon sensor

M. H. Abdel Basset, Eman Yossri Frag, M. A. Zayed
Scientific Reports
Analytical Chemistry and Sensors
article

Quantitative determination of al (III) traces in soft drink, pharmaceutical products and biological fluids of kidney failure and alzheimer disease patients using carbon sensor

M. H. Abdel Basset, Eman Yossri Frag, M. A. Zayed
article en

Abstract

Carbon paste sensor based on p-chlorophenyl maleanilic acid (pCl-MA) as an ionophore was synthesized and investigated as an Al(III) ion selective sensor. The developed sensor shows Nernstian slope of 20.32 ± 1.18 mV decade− 1 over the working concentration range of 1.0 × 10− 6 − 1.0 × 10− 1 mol L− 1 of Al (III) ion with a detection limit of 3.3 × 10− 7 mol L− 1. The sensor exhibited a stable potential response in the pH range from 4.0 to 6.0 and a fast response time of 8 s. The prepared sensor indicated a very good selectivity to Al (III) over a variety of cations. The sensor could be easily employed for determination of Al (III) ions in pharmaceutical products, soft drink and biological fluids of kidney failure and Alzheimer disease patients. The reaction mechanism between pCL-MA and Al (III) was confirmed by FT-IR spectroscopic analysis, SEM and EDX.

Scientific ReportsVol. 16(1)
Cairo University (EG)
Science and Technology Development Fund
Openalex Percentile: Top 96%
Analytical Chemistry and Sensors
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.