Non-Invasive Blood Pressure Estimation Enhanced by Capillary Refill Time Modulation of PPG Signals
This study evaluates the impact of capillary refill time (CRT) modulation on photoplethysmography (PPG) signals for improved non-invasive continuous blood pressure (CBP) estimation. Data from 21 healthy participants were collected, applying a standardized 9 N pressure for 15 s to induce CRT during 6-min sessions. PPG signals were segmented into 252 paired 30-s intervals (CRT-modulated and standard). Three machine learning models-ResNetCNN, LSTM, and Transformer-were validated using leave-one-subject-out (LOSO) and non-LOSO methods. CRT modulation significantly enhanced accuracy across all models. ResNetCNN showed substantial improvements, reducing mean absolute error (MAE) by up to 35.6% and mean absolute percentage error (MAPE) by up to 40.6%. LSTM and Transformer models also achieved notable accuracy gains. All models met the Association for the Advancement of Medical Instrumentation (AAMI) criteria (mean error < 5 mmHg; standard deviation < 8 mmHg). The findings suggest CRT modulation's strong potential to improve wearable CBP monitoring, especially in resource-limited settings.